Страница 2: ...2 1 2 3 4 Fig 2 Fig 4 Fig 8 Fig 5 Fig 6 K L M T U Q S U Fig 1 Q B M O N C L K A J I H G F E P D EUROSTAR station Wireless Controller WiCo P RS 232 USB Temp Sensor Fig 7 X I G Z H G W V Fig 3 H I J G G ...

Страница 3: ...rference received including interference that may cause undesired operation NOTE This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can r...

Страница 4: ...ce should ensure that RF radiation is not emitted in excess of the Health Canada s requirement Explication of warning symbols Indicates an extreme hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can result in death serious injury Indicates a potentially hazardous situation which if not avoided can result in injury In...

Страница 5: ...serve the permitted speed for the stirring ele ment Never set higher speed Make certain that the unit is set at the lowest speed be fore commissioning otherwise the unit will begin run ning at the speed set in last operation Gradually increase the speed Pay attention when setting the speed to any imbalance of the stirrer tools and possible spraying of the medium to be stirred Never operate the dev...

Страница 6: ...er WiCo is fully discharged the only means of switching the station off are the safe button and the off switch Please note the following safety instructions for the battery pack RB 1 recharge able battery Keep the battery pack out of reach of children at all times Store the battery pack in a cool dry place Never throw the battery pack into a fire Keep it away from direct sunlight and temperatures a...

Страница 7: ...be performed using a USB cable This device is suitable for use in all areas except Residential areas Areas that are connected directly to a low voltage supply network that also supplies residential areas The safety of the user cannot be guaranteed If the device is operated with accessories that are not sup plied or recommended by the manufacturer If the device is operated improperly contrary to th...

Страница 8: ...ing IKA recommend stirrer tools see section Permitted IKA Stirrer tools The output shaft is designed as a hollow shaft and the opening on the top side of the housing is closed with a plastic cover When the stirrer cover is opened it is possible for stirring shaft to be pushed out over the top edge of the housing at standstill e g during changing container For safety reason the stirrer cover must b...

Страница 9: ... 5 Use a stirring shaft protector Q e g R 301 to provide pro tection against injury when working with the device Use the bolts U to attach the plastic half shell pieces to the stirrer T as shown in Fig 5 The screw S can be used to adjust the length of the stirring shaft protector Check that the stirring shaft protector is held in position securely prior to each use and also at regular intervals Th...

Страница 10: ... second time enables the controls again key symbol disappears from the display In emergency the stirrer station can be switched off by pressing the safe Stop key I see Fig 1 on the front of the stirrer In this case the LED bar G see Fig 1 changes color from green to red and flashes A mes sage appears in the display advising that the EUROSTAR stirrer station has been forcibly switched off To restart...

Страница 11: ...roller on and off L Lock key Key disable and rotary push knob disable N Menu key Press it once main menu is displayed Press it a second time back to the working screen B Rotary push knob Navigation selecting and changing the settings in the menu O Back key Return to the previous menu level reset the torque K Gear key Change the gear in different speed torque ranges Note Can only change the gear in...

Страница 12: ... you press the Back key O the system skips back to the previous display Back Back Back Menu Menu Menu Tempe Stirring Torque Calibration Stirring Intermittent Mode Torque Display Torque Limit Speed Limit h Stirring Speed Limit Speed Limit OK 1 5 0 0 rpm B 0 0 actual rpm set rpm Torque 0 Ncm Reset Torque J Back Timer 00 00 00 Press the menu key N Select the menu by turning the rotary push knob B to ...

Страница 13: ...tivated Safety Time Out yes Safe Speed yes yes Information Stirring Max Speed Max Torque yes Interval Run yes Interval Stop yes Software Version yes Operating Mode yes Station Version yes Firmware Update Info yes Units Temperature ºC activated ºF Display Background Black activated White Firmware Update Info Brightness Standard Mode 60 Battery Mode 20 Key Tone Sound Volume 100 Bluetooth activated R...

Страница 14: ...the value specified as the maximum torque limit Note The torque limit can be overload for about 10 sec onds This is necessary to cater for stirring tasks which re quire metering and feeding additives Torque Display The Torque Display menu allows the user to specify that the torque is shown in the display A tick shows that the option is activated Note Pressing the Back key while the device is runnin...

Страница 15: ...ears in front of the Ncm unit Programs Under menu Programs 5 user defined speed rpm time profiles can be created In addition the user could define the intermittent mode is activated or not in the programs If the intermittent mode is activated the run time stop time value from Intermittent Mode setting will be taken Note If user need to activate the intermittent mode in one segment of a program he sho...

Страница 16: ... Mode No PROGRAM 1 Detail Press the rotary push knob on option Detail the detailed settings of every segment can be seen Turn the rotary push knob the details screen of every seg ment can be switched Insert With the insert option a new segment will be insert below the selected segment The default settings for the new seg ment are as following rpm 0 Time hh mm 00 01 Intermittent Mode No View the pr...

Страница 17: ...nterm Mode Edit Delete Insert 1 2 3 4 5 150 0 00 10 00 00 Yes No Programs Start Edit Delete Program 1 Program 2 Program 3 Program 4 Program 5 PROGRAM 1 Seg No rpm Time hh mm Interm Mode 1 2 3 4 5 0 00 00 No Edit Delete Insert Back Programs Start Edit Delete Program 1 Program 2 Program 3 Program 4 Program 5 Programs Edit Delete Program 1 Program 2 Program 3 Program 4 Program 5 Start Programs Edit D...

Страница 18: ... from un foreseen vibration When the set vibration level is reached the device will shut off automatically Level 0 no vibration protection Level 1 low vibration limitation Level 2 middle vibration limitation Level 3 high vibration limitation Settings Language The Language option allows the user to select the desired language by turning and pressing the rotary push knob B A tick indicates the langu...

Страница 19: ...s 1 parity bit even 1 stop bit Transmission speed 9600 bit s Data flow control none Access procedure data transfer from the stirrer machine to the computer takes place only at the computer s request Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the stirrer machine Slave The stirrer machine sends only at the computer s requ...

Страница 20: ...puter A USB micro B USB A 9 pin RS 232 9 pin RS 232 Oder micro B USB A USB A USB A A USB micro B USB A micro B Oder 9 pin RS 232 9 pin RS 232 A USB micro B USB A micro B PC 1 1 Cable station to PC Required for connecting the 9 pin socket to a PC 1 2 RxD 3TxD 4 5 GND 6 7 RTS 8 CTS 9 1 RxD 2 TxD 3 4 GND 5 6 RTS 7 CTS 8 9 PC 1 2 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 ...

Страница 21: ...uest the Decontamination Certifi cate form IKA or use the download printout of it from IKA website www ika com If you require servicing return the device in its original pack aging Storage packaging is not sufficient Please also use suitable transport packaging Error codes The fault is shown by an error code in the display C as following if the error occurs e g Error 4 Proceed as follows in such ca...

Страница 22: ...glass coated See more accessories on www ika com Permitted IKA stirrer tools R 1342 Propeller stirrer 2000 R 1345 Propeller stirrer 800 R 1381 Propeller stirrer 2000 R 1382 Propeller stirrer 2000 R 1385 Propeller stirrer 800 R 1388 Propeller stirrer 400 R 1389 Propeller stirrer PTFE 800 R 2302 Propeller stirrer 600 R 1311 Turbine stirrer 2000 R 1312 Turbine stirrer 2000 R 1313 Turbine stirrer 800 ...

Страница 23: ...age VAC 230 10 115 10 100 10 Frequence Hz 50 60 Max input power W 226 Max output power at stirring shaft W 167 Protection class according to EN 60529 IP 40 Protection class I Excess voltage category II Contamination level 2 Protection at overload yes motor current limitation Fuse on mains plate A T 6 3 A IKA Ident No 2585000 Drive Brushless motor Ambient temperature C 5 40 Ambient humidity rel 80 ...

Страница 24: ...erance PT 1000 DIN EN 60751 class A max communication range dependent on the building m 40 150 Dimension W x D x H WiCo mm 71 x 74 x 151 Weight WiCo kg 0 28 USB interface WiCo yes RB 1 Battery pack Voltage V 3 7 Battery capacity mAh 2000 Charging time h 4 5 Working time h 15 Battery type Lithium polymer Subject to technical changes ...