13
In case the excess steam turns into water inside the oven (chamber), it may be drained
through the water outlet pipe located between the bottom of the chamber and the oven burners. It is
advisable to install a point to drain this water, avoiding that the floor is wet and slippery and can
cause accidents.
3.3 Electrical Connection
The equipment is supplied with a plug-in power cable to be connected directly to the mains
or to an industrial type plug suitable for each type of current (refer to the current for each model of
equipment in the table of item 2.2 Technical Data).
The power cable for the oven model with 220 V single-phase connection has three wires
(Ground, Phase and Neutral), the model with voltage 220 V three-phase connection with four wires
(Ground and Phases RST) and voltage 380 V three-phase connection has 5 wires (Ground, Neutral
and RST Phases).
For cables that have Neutral, their color is blue and is labeled with a sticker. It is mandatory to
connect the neutral wire to the neutral of the mains.
If the power cord is damaged, it must be replaced with a new one. The exchange must be made
by the manufacturer, authorized agent or a qualified person in order to avoid risk.
ATTENTION!
In order to avoid accidents, it is mandatory to connect the neutral wire to the neutral of the
power grid. And also the ground wire for all types of voltage, according to the current regulation.
Never connect the ground wire to the neutral wire of the mains, water pipes, gas pipes, etc.
For correct grounding, consider the instructions in standard NBR 5410 - ABNT.
Accompanying the equipment, an equipotential grounding terminal (Electric Terminal), located on
the right-hand side of the oven
.
Содержание BIMG METVISA FTE.150
Страница 1: ...1 ...
Страница 9: ...9 Follow the guidelines below to assemble the stand ...
Страница 26: ...27 Electrical Schematic FTE 300 Voltage 220 V Three Phase Connection ...
Страница 28: ...29 Electrical Schematic FTE 300 Voltage 380 V Three Phase Connection ...
Страница 29: ...30 Explosion Design ...
Страница 30: ...31 Explosion Design ...
Страница 31: ...32 Explosion Design ...
Страница 32: ...33 Explosion Design ...
Страница 38: ...39 Notes ...
Страница 39: ......