D
x
Telemetry System
Manual Vers. 2.7
37
/
92
3.8
Configure Analog Output
Any measurement data channel of the D
X
can be assigned to one of the 6 analog outputs.
Select
Output → Output X
.
Figure 26: Analog output configuration
With
Source
, select the data channel to be displayed.
Check the
min/max = -10 V/+10 V
box to set the upper/lower bounds of the measurement
range (see 3.6.4.3) to -10 V/+10 Volts. If the box is unchecked, the measurement range can
be defined by a two-point calibration.
Please note:
Maximum/minimum analog output is ± 10 Volts.
When configuring through a web browser, confirm your input with the
Set
button to resume it
to the D
X
.
Содержание CAEMAX DX
Страница 31: ...Dx Telemetry System Manual Vers 2 7 31 92 Figure 20 Strain gauge channel configuration...
Страница 44: ...Dx Telemetry System Manual Vers 2 7 44 92 Menu tree continued...
Страница 84: ...Dx Telemetry System Manual Vers 2 7 84 92 10 Questions Answers 10 1 Inductive Power Supply Troubleshooting...
Страница 85: ...Dx Telemetry System Manual Vers 2 7 85 92...
Страница 86: ...Dx Telemetry System Manual Vers 2 7 86 92...