Description
3 Description
3.1
Design and Function
3.1.1 Laboratory-Vacuum-Systems
The Laboratory Vacuum Systems are supplied as a ready-mounted unit.
The chemical-resistant diaphragm pump
(4)
is fitted to the base plate of the column
(3)
and
attached on the suction-side to the separator
(7)
to protect the pump from foreign matter and
on the pressure-side to the KD 500/5 emission condenser
(6)
to recover the condensate.
This ensures that hazardous substances are separated out. The safety valve on the emis-
sion condenser protects against an impermissibly high operating pressure on the outgoing
air end.
The complete vacuum controller, with sensor, venting valve and power unit
(5)
, is fitted in the
column
(3)
. (The non-controlled LVS-variant does not have a controller.)
All exposed glass components
(6, 7)
are coated with transparent plastic as protection a-
gainst bursting. The emission condenser is supplied with insulation.
The power switch
(12)
and the plug for the power cable
(11)
as well as the membrane pump
(10)
, the PC
(13)
control lead
(14)
and the water valve (option)
(15)
are located on the right
of the column.
The LVS types are specially preconfigured according to model.
1
Pressure side connection
hose nozzle for hose DN8
2
Control valve
(controlled in the case of the LVS)
3
Column complete
on swivelling foot
4
Diaphragm pump MPC 101 Z
with gas ballast valve and main switch
5
Vacuum controller 521
6
Emission condenser, short
KD 500/5 insulated
at the pressure side with safety valve
and cooling water connections
7
Separator
(round-bottomed flask) suction side
Fig. 1a LVS 110 Z - Front view
8
Cooling water
(feed flow/back flow)
9
Connection outgoing air
hose nozzle for hose DN8
10
Connection Pump
(pump)
11
Connection non-heating device power
cable
(main in)
12
Main switch
(power)
13
Connection PC
(RS 232)
14
Connection control cable
controlled LVS (sensor)
15
Connection water valve
16
Connection inert gas
17
Gas ballast valve
Fig. 1b LVS 110 Z - Rear view
10 Ilmvac
GmbH
115022