Item
Storage
Instructions
XC2
-25°C to -15°C
Thaw at room temperature. Invert 10 times to mix.
XC3
Room
temperature
Thaw at room temperature. Invert 10 times to mix.
XC4
Room
temperature
Thaw at room temperature. Invert 10 times to mix.
Procedure
Wash BeadChips
1
Attach the wire handle and submerge the wash rack in Wash 1 containing 200 ml PB1.
NOTE
Replace PB1 in Wash 1 after every 12 BeadChips.
2
Remove the hybridization chamber inserts from the hybridization chambers.
3
Remove the BeadChips from the hybridization inserts.
4
Remove the cover seals from the BeadChips.
Using powder-free gloved hands, hold the BeadChip securely and by the edges in one hand. Remove
the entire seal in a single, continuous motion. Do not touch exposed arrays.
5
Immediately and carefully slide each BeadChip into the wash rack in Wash 1, making sure that the
BeadChip is submerged in PB1 a maximum of 8 BeadChips)..
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
21
Infinium HD Super Assay Reference Guide