3
Wipe the surface of the flow cell holder using an alcohol wipe or a lint-free tissue moistened with ethanol
or isopropanol.
CAUTION
Do not allow alcohol to drip into the vacuum holes or around the manifolds. Use a low-lint lab tissue to
dry the stage, if necessary.
4
Load 10 ml laboratory-grade water in each position in the reagent racks, and then lower the sippers.
5
Turn off the instrument.
6
To restart the instrument:
a
Load water in all reagent positions.
b
Turn on the instrument.
c
Perform a water wash.
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
48
HiSeq 2000 System Guide