a
Pull the handle towards you and raise it.
b
Release the handle into the slot on the top end of the groove. Make sure that the handle rests
securely in the slot.
3
Slide the reagent rack out of the reagent compartment using the rack handle.
4
Remove the cap from each reagent tube.
5
Place each tube onto the rack in the associated numbered positions.
Position
Reagent
Description
10
RMR
Resynthesis Mix
11
LMX2
Linearization Mix 2
12
BMX
Blocking Mix
13
AMX2
Amplification Mix 2
14
APM2
AMX2 Premix
15
AT2
100% Formamide
16
HP7 or HP11
Read 2 Sequencing Primer
18
HP3
Denaturation Solution
19
HT2
Wash Buffer
6
Slide the reagent rack into the reagent compartment.Align the rack with the raised guide on the floor of
the compartment.
7
Lower the sippers into the paired-end reagent tubes as follows.
a
Pull the handle towards you and lower it.
b
Inspect the sippers to make sure that they do not bend as they lower into the tubes.
c
Release the handle into the slot on the bottom end of the groove.
Load ICB for Read 2
1
Raise the sippers for the SBS reagent rack as follows.
a
Pull the sipper handle towards you and then raise it.
b
Release the handle into the slot on the top end of the groove. Make sure that the handle rests
securely in the slot.
2
Slide the reagent rack out of the reagent compartment.
3
Remove the ICB reagent bottle from position 1 and remove the funnel cap.
4
Place the funnel cap on the new bottle of ICB and load the bottle into position 1. Make sure that the
conical end of the bottle rests in the indentation on the base of the rack.
5
Slide the reagent rack into the reagent compartment. Align the rack with the raised guide on the floor of
the compartment.
6
Lower the sippers into the SBS reagent bottles as follows.
a
Pull the sipper handle towards you and then lower it.
b
Inspect the sippers to make sure that they do not bend as they lower into the funnel caps.
c
Release the handle into the slot on the bottom end of the groove.
7
Close the reagent compartment door.
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
42