A first base report is generated automatically for each run. Selecting this option opens the first base
report before the run proceeds.
2
[Optional]
Clear the Align to PhiX checkbox of any lane that does not contain PhiX.
Alternatively, select lanes on the flow cell image to add or remove lanes for PhiX alignment.
By default, all lanes are selected for alignment.
NOTE
With HCS v2.2 and RTA v1.18, a dedicated control lane is not required or an option.
3
Select Next.
Recipe Screen
1
Select from the following Index Type options:
u
No Index—Performs a non-indexed single-read or paired-end run.
u
Single Index—Performs a single-read or paired-end run with 1 indexing read.
u
Dual Index—Performs a single-read or paired-end run with 2 indexing reads.
u
Custom—Performs a single-read or paired-end run with a custom number of cycles for index reads.
2
If the Dual Index or Custom option is specified, select a Flow Cell Format, either Single Read or Paired
End.
3
Enter the number of cycles for Read 1 and Read 2, if applicable.
The number of cycles performed in a read is 1 more cycle than the number of cycles analyzed. For
example, to perform 125 cycles for Read 1, enter 126.
4
For the Custom indexing option, enter the number of cycles for index reads.
NOTE
Read lengths do not need to be identical.
5
Confirm the following autopopulated chemistry settings:
u
SBS: HiSeq SBS Kit v4
u
Index: HiSeq v4 Single Index or HiSeq v4 Dual Index
u
PE turnaround: HiSeq PE Cluster Kit v4
6
[Optional]
Select the Use Existing Recipe checkbox to use a custom recipe.
Sample Sheet Screen
Sample sheets are optional unless you use BaseSpace Sequence Hub to perform data analysis or perform an
indexed run.
NOTE
HCS v2.2 allows a different indexing scheme in each lane.
1
Select Browse to locate the sample sheet.
2
Select Next.
Reagents Screen
The Reagents screen records information about reagent kits used for the run. The reagent kit ID (barcode
number beginning with RGT) determines reagent kit type and run mode compatibility.
1
Scan or enter the SBS reagent kit ID.
Document # 15011190 v03
For Research Use Only. Not for use in diagnostic procedures.
15
HiSeq 2000 System Guide