cBot 2 System Guide
65
Change the Adapter Plate
You can use a HiSeq flow cell or a GAIIx flow cell on the cBot. Each flow cell type requires
that a specific adapter plate is installed. Icons on the Start screen indicate which adapter
plate is installed.
NOTE
The cBot is shipped with the HiSeq adapter plate installed.
1
Open the instrument lid by gently lifting from the cutout on the front of the lid.
2
Lift the flow cell clamp.
3
Loosen the 2 captive Phillips head screws securing the adapter plate.
Figure 29
Flow Cell Adapter Plate
A
Captive screws
B
Adapter plate
4
Lift the existing adapter plate from the thermal stage and set aside.
5
If salts are present on the thermal stage, wipe with a lint-free cleaning tissue slightly
moistened with water.
6
Position the new adapter plate on the thermal stage. Align the sensor arm with the
corresponding slot on the right side of the thermal stage.
Figure 30
Sensor Arm Location
A
GAIIx adapter plate
B
HiSeq adapter plate
C
Adapter plate sensor arm
7
Tighten the 2 screws to secure the adapter plate. For optimal heat transfer, make sure
that the adapter plate is sitting flat and the screws are tightened evenly.
Содержание cBot
Страница 4: ...iv Material 20004364 Document 15065681 v01 ...
Страница 16: ...10 Material 20004364 Document 15065681 v01 ...
Страница 22: ...16 Material 20004364 Document 15065681 v01 ...
Страница 66: ...60 Material 20004364 Document 15065681 v01 ...
Страница 76: ...70 Material 20004364 Document 15065681 v01 ...
Страница 88: ...Index 82 Material 20004364 Document 15065681 v01 ...