6 - Operation
Operating in Routine Operation Mode
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
6 - 9
6.3.4 Instrument Preparation and Checks
Cross
References
List of cross references to information provided in other sections:
General
Information
This section contains instructions for routine use. It is intended as a guide to build
your SOP (Standard Operating Procedure).
Any modifications of the implemented tests in your Application Software must be
carried out by application specialists or expert operators.
Refer to cross references above.
Before starting a run, pay attention to the following:
Containers
1
Empty the waste liquid container if necessary.
The waste container must remain on floor level to provide for proper waste
liquid flow.
2
Empty the disposable tip waste bag if necessary.
3
Check the system liquid container and refill if necessary.
If possible, place the system liquid container on worktable level to avoid
pressure difference in the supply tubing.
Consumables
4
Check the disposable tip rack and add tips if necessary.
5
Make sure that the reagent troughs are filled appropriately.
6
Ensure that the daily maintenance has been carried out according to the
maintenance chapter.
Furthermore, take the following notices into consideration:
Worktable
Regarding the worktable, pay attention to the following:
ATTENTION
Improper positioning of objects on the worktable may lead to disturbances or
errors in the process, e.g. misinterpretation of barcodes. Do not use free space on
the worktable to deposit any objects
Subject
Reference
User qualification
See section
No air bubbles in the tubing
See section
7.3.1.2 “Flushing the Liquid Sys-
No liquid droplets on tips
See section