DIGIOHMpro
- 17 -
5.2. General data
Power supply
4x AAA alkaline battery 1,5 V or NiMH accumulator 1,2 V
Over voltage class:
- protective caps of the test tips inserted
CAT III 300 V
- protective cap(s) of the test tip(s) removed CAT II 300 V
Pollution degree
2
Protective class
II (double insulation)
Degree of protection
IP 40
Dimensions
about 255x70x40 mm
Weight including batteries and movable test tip
about 0,36 kg
Maximum altitude for use of the device
2000 meters
Reference conditions
ambient temperature (23 ± 2) °C
relative humidity 40 ÷ 60 % (non condensing)
mains voltage 230 V ± 2 % / 50 Hz ± 1 %
instrument’s position arbitrary
Operating conditions
ambient temperature 0 ÷ 40 °C
relative humidity max. 85 % (non condensing)
mains voltage 190 ÷ 260 V / 48 ÷ 52 Hz
instrument’s position arbitrary
Storage conditions
ambient temperature -10 ÷ +70 °C
relative humidity
max. 90 % (-10 ÷ 40) °C
(non condensing)
max. 80 % (40 ÷ 70) °C
instrument’s position arbitrary
Note:
The instructions for use of the instrument in this manual are current when
delivered together with the instrument, but may not match the older or newer
version of the instrument’s firmware.
Firmware can also be updated during calibration and repair of the instrument.
Updating the firmware can cause changes and / or additions to the device
functions so we recommend checking the validity of the manual on www.illko.cz
in relation to the firmware version.