IKONIX SCI 290 Series Скачать руководство пользователя страница 1

Operation & 

Service Manual

Model 294, 295, 296, 297, 298 • V1.05

Model 294 (DC Hipot Tester)

Model 295 (AC Hipot Tester)

Model 296 (AC/DC Hipot Tester)

Model 297 (AC/DC Hipot With IR Tester)

Model 298 (500VA AC Hipot Tester)

Item 99-10813-01

Printed May 21, 2020

RoHS 2

COMPLIANT

2011/65/EU

Содержание SCI 290 Series

Страница 1: ...95 296 297 298 V1 05 Model 294 DC Hipot Tester Model 295 AC Hipot Tester Model 296 AC DC Hipot Tester Model 297 AC DC Hipot With IR Tester Model 298 500VA AC Hipot Tester Item 99 10813 01 Printed May...

Страница 2: ...4 2012 IEC 61000 4 4 2012 EN 61000 4 5 2006 IEC 61000 4 5 2005 EN 61000 4 6 2014 IEC 61000 4 6 2013 EN 61000 4 8 2010 IEC 61000 4 8 2009 EN 61000 4 11 2004 IEC 61000 4 11 2004 Supplementary Informatio...

Страница 3: ...014 IEC 61000 4 5 2014 EN 61000 4 6 2014 IEC 61000 4 6 2013 EN 61000 4 8 2010 IEC 61000 4 8 2009 EN 61000 4 11 2004 IEC 61000 4 11 2004 Supplementary Information The product herewith complies with the...

Страница 4: ...return of the tester for warranty service must be prepaid by the customer SCI will assume the return freight costs when returning the tester to the customer The return method will be at the discretion...

Страница 5: ...ls On Remote I O 33 Using The Tester Accessories 34 Using The Test Leads 34 Using The Adapter Box 35 Appendix A Installation And Test Operator Information 36 Installation 36 1 Unpacking And Inspection...

Страница 6: ...NSTRUCTION MANUAL FOR SPECIFIC WARNING OR CAUTION INFORMATION TO AVOID PERSONAL INJURY OR DAMAGE TO THE PRODUCT INDICATES HAZARDOUS VOLTAGES MAY BE PRESENT CHASSIS GROUND SYMBOL WARNING CALLS ATTENTIO...

Страница 7: ...nd Return ports LCD DISPLAY The Liquid Crystal Display is the main readout for the operator and programmer of the test settings and test results HIGH VOLTAGE OUTPUT JACK Use this jack for the connecti...

Страница 8: ...IONAL CONTINUITY CHECK PORT Provides the connection for checking ground continuity OPTIONAL RETURN PORT Provides the return connection for the leakage current OPTIONAL HIGH VOLTAGE OUTPUT PORT Use thi...

Страница 9: ...the button will light when there is continuity between Cont Check and Return ports LCD DISPLAY The Liquid Crystal Display is the main readout for the operator and programmer of the test settings and...

Страница 10: ...mmunication Refer to Option 03 in the Options section SIGNAL OUTPUT 9 pin D subminiature female connector for monitoring PASS FAIL and PROCESSING output relay signals SIGNAL INPUT 9 pin D subminiature...

Страница 11: ...e of the front panel Upon powering the tester up a POWER ON SELF TEST POST will automatically be performed This test will check for the condition of the RAM chips PCB s and other critical components I...

Страница 12: ...ts loaded in each memory location can be executed one at a time or as a sequence 2 Perform Test Screen The Perform Test Screen displays the following parameters Follow the setup procedures to set the...

Страница 13: ...ecting Memory START GO BACK TO THE MAIN MENU Turn the yellow rotary knob to scroll to BACK in order to return to the Main Menu Main Menu When BACK is selected Press the rotary knob You will be returne...

Страница 14: ...Press the rotary knob to select the Test Type You will be returned to the previous menu START PROGRAM A MEMORY CHANGE VOLTAGE When VOLT is selected Press the rotary knob to edit the Voltage The first...

Страница 15: ...knob to select the setting and move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE LO LIMIT When LO L is selected Press the rotary knob to edit th...

Страница 16: ...nob to select the setting and move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE DWELL When DWLL is selected Press the rotary knob to edit the Dwe...

Страница 17: ...er a value for HI Limit Then enter a value for LO Limit Then enter a value for Offset Then enter an offset value manually Then press the test button Short the Cont Check lead with the Black Return Lea...

Страница 18: ...t setting ON or OFF Press the rotary knob to select the setting You will be returned to the previous menu START PROGRAM A MEMORY CHANGE FREQUENCY When FREQ is selected Press the rotary knob to edit th...

Страница 19: ...to the previous menu START PROGRAM A MEMORY TYPE DC When TYPE is selected Press the rotary knob to edit the Test Type The options will blink Turn the yellow rotary knob to scroll to desired parameter...

Страница 20: ...knob to select the settingand move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE HI LIMIT When HI L is selected Press the rotary knob to edit the...

Страница 21: ...y knob to select the setting and move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE RAMP When RAMP is selected Press the rotary knob to edit the R...

Страница 22: ...er a value for HI Limit Then enter a value for LO Limit Then enter a value for Offset Then enter an offset value manually Then press the test button Short the Cont Check lead with the Black Return Lea...

Страница 23: ...u START PROGRAM A MEMORY TYPE IR When TYPE is selected Press the rotary knob to edit the Test Type The options will blink Turn the yellow rotary knob to scroll to desired parameter to edit TYPE Turn t...

Страница 24: ...to select the setting and move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE HI LIMIT IR When HI L is selected Press the rotary knob to edit the H...

Страница 25: ...tary knob to select the setting and move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE RAMP IR When RAMP is selected Press the rotary knob to edit...

Страница 26: ...ch digit Press the rotary knob to select the setting and move to the next digit When done you will be returned to the previous menu START PROGRAM A MEMORY CHANGE CONNECT When CONN is selected Press th...

Страница 27: ...stem parameter the following message will be displayed Security is RUN can t be edited MEM With SCTY set to MEM the user will be able to recall any of the 5 available memory locations The user will no...

Страница 28: ...tings The options will blink Turn rotary knob to change the Security settings OFF RUN or MEM Press the rotary knob to select the Security setting option Enter a PIN number for the Security setting Tur...

Страница 29: ...OFF Press the rotary knob to select the PLC setting You will be returned to the previous menu START PROGRAM A MEMORY CHANGE DUAL When DUAL is selected Press the rotary knob to edit the Connect Settin...

Страница 30: ...Hipot Displayed if the leakage current exceeds the high limit setting but does not exceed the metering range Using the Display Test Mode Displays Test Mode Display Test Type Description AC DC Hipot I...

Страница 31: ...y Description This message appears on the display if the tester s output reading does not match the setting When the tester has an output problem and the TEST button is pressed the Output Error screen...

Страница 32: ...e 745171 1 Shielded Cable Clamp for either Plug or Receptacle 747784 3 Jackscrew Set 2 Reviewing Test Results for Multi step Sequences After the test is performed the test results will be indicated on...

Страница 33: ...on For safety the front panel RESET but ton remains active even when a remote reset switch is connected so that high voltage can be shut down from either location INTERLOCK 4 and 5 Remote Interlock ut...

Страница 34: ...ester Plug the continuity lead to the CONT CHECK terminal on the front panel of the tester Connect the clip end continuity lead to the ground pin of the DUT Connect the clip end of the high voltage le...

Страница 35: ...box to the front panel CONT CHECK terminal Connect the high voltage lead from the adapter box to the front panel high voltage terminal Connect the black return lead 99 10008 01 to the front panel retu...

Страница 36: ...er Determine if the tester can be lifted by one individual or requires additional support Make sure that your balance is centered and your feet are properly spaced shoulder width apart behind the test...

Страница 37: ...ection provided by the tester may be impaired WARNING BEFORE CONNECTING POWER TO THIS TESTER THE PROTECTIVE GROUND EARTH TERMINALS OF THIS TESTER MUST BE CONNECTED TO THE PROTECTIVE CONDUCTOR OF THE L...

Страница 38: ...e container FRAGILE to insure proper handling Please refer to the RMA number for all correspondence Packaging Instructions 1 Be sure to REMOVE ALL ACCESSORIES and the INTERLOCK DISABLE from the tester...

Страница 39: ...ectrodes a component tested with the high voltage test lead or a cord connected device with a two or three prong plug Plug in the high voltage test lead only when it is being used Handle its clip only...

Страница 40: ...r should be easily accessible Keep the area clean and uncluttered All test equipment and test leads not absolutely necessary for the test should be removed from the test bench and put away It should b...

Страница 41: ...ng 10 V 500 V 1 5 of reading 500 V Current Display Range Resolution Accuracy 0 10 12 00 mA AC 0 02 5 00 mA DC 0 01 mA 2 of reading 2 counts Model 298 Range Resolution Accuracy 0 10 100 mA AC 0 01 mA 2...

Страница 42: ...tput Frequency Range Accuracy 60 or 50 Hz User Selectable 0 1 Output Regulation 1 of setting 5 V from no load to full load Dwell Timer Range Resolution Accuracy 0 0 2 60 0 sec 0 Continuous 0 1 sec 0 1...

Страница 43: ...software commands for maximum current measured during the test including Ramp Up and Dwell Time Line Cord Detachable 6 ft 1 8 m power cable terminated in a three prong grounding plug Terminations 6 ft...

Страница 44: ...ear panel of the unit when options are present Model Option Label On the rear panel of the tester you will find a label that contains the option code For example your options code will appear as follo...

Страница 45: ...for safety when this lead is not being used This jack is always active when performing a test Option 03 USB interface This option may be added as a serial type communication protocol This option provi...

Страница 46: ...racy 0 00 3 00 0 01 2 of setting 2 counts Measurement AC Current mA Range Resolution Accuracy 0 10 3 00 0 01 2 of reading 2 counts Short Circuit Protection minimum current 3 5 mA RMS at short circuit...

Страница 47: ...nuous testing The testing time will not be restarted when automatic reset for continuous testing With this option installed on the tester an additional test parameter PULS will be available in the men...

Страница 48: ...options for the parameter ON or OFF You will then be returned to the previous menu Confirm by pressing the rotary knob Option 08 Push to Test Mode With this option installed an extra parameter called...

Страница 49: ...data flow When requesting data from the tester it will automatically send the data back to the controller input buffer The controller input buffer will accumulate data being sent from the tester incl...

Страница 50: ...er 1 5 FL Query Memory Location 1 5 SAA Add ACW Test SAD Add DCW Test SAI Add IR Test ADD ACW p p p p p p p p p p p Add all parameters for ACW Test ACW Voltage kV HI Limit mA LO Limit mA Ramp s Dwell...

Страница 51: ...ferent possible set of values Command Name Test Types Values EV value EV Edit Voltage ACW DCW IR 0 00 5 00 kV 0 00 6 00 kV 100 1000 V EH value EH Edit HI Limit ACW DCW IR 0 10 12 00 mA 0 02 5 00 mA 0...

Страница 52: ...eter is separated by commas and includes memory location test type test status and metering The syntax for this command response is memory step test type status meter 1 meter 2 meter 3 Memory GND Stat...

Страница 53: ...ual step indicated by step number 1 5 The response will be formatted as follows SFW Return the current firmware version of the tester IEEE 488 2 Common Commands These commands are required by the IEEE...

Страница 54: ...status clear flag TST Performs a self test of the tester data memory Returns 0 if it is successful or 1 if the test fails CLS Clears the Status Byte summary register and event registers Does not clear...

Страница 55: ...the Status Byte Returns the decimal value of the binary weighted sum of bits SRE value Service Request enable register controls which bits from the Status Byte should be use to generate a service requ...

Страница 56: ...Return Connector 99 10312 01 1 Power Switch 2P 15A 99 10016 01 1 Earth Connector 99 10735 01 1 Rotary Knob 330 113 001 1 TEST Test Switch Green Lighted 330 113 002 1 RESET Reset Switch Red Lighted 99...

Страница 57: ...c Refer servicing to a SCI authorized service center Service Interval The tester and its power cord test leads and accessories must be returned at least once a year to a SCI authorized service center...

Страница 58: ...with 6 000 Volts range A Standard DC Voltmeter with 1 000 Volts range A Standard AC Milliammeter with 20 mA range A Standard DC Milliammeter with 5 mA range A Standard Resistor 100 k 20Watt 1500VAC A...

Страница 59: ...Voltage Please connect a standard 6000V DC Voltmeter to the HV and RETURN connectors Rotate the yellow knob until the display shows Then press the TEST button on the front panel The tester will provi...

Страница 60: ...begin outputting high voltage on the output connectors Then the display will show a value similar to that of the standard Turn and press the rotary knob to enter the reading of the standard AC Ammeter...

Страница 61: ...rd AC Ammeter into the tester Once all digits are entered the tester will advance to the next calibration point Then press the Test button on the front panel The tester will begin outputting high volt...

Страница 62: ...anel The tester will execute automatic calibration process The process does not require data entry The tester will provide around 300VDC on the output connectors and the display will show 7 To calibra...