IKM TC15 Скачать руководство пользователя страница 1

User manual

TC65/TC15 Temperature Calibrator

Содержание TC15

Страница 1: ...User manual TC65 TC15 Temperature Calibrator ...

Страница 2: ...ic mechanical photocopying recording or otherwise without the prior permission of IKM Instrutek AS All efforts have been made to ensure the accuracy of this handbook IKM Instrutek is always striving to improve our products and handbooks We would greatly appreciate being informed of any errors found in our product or in its manual The above notwithstanding IKM Instrutek cannot take responsibility f...

Страница 3: ...hy calibrate 5 How often do I calibrate 6 How to use the calibrators 7 Safety troubleshooting and maintenance 8 Technical information 9 Appendix p 5 p 8 p 9 p 10 p 11 p 12 p 15 p 16 p 17 EC Declaration of Conformity according EN ISO 17050 1 2010 is supplied with the product EMC test and approval are performed by Nemko according to EN61326 1 2013 Type approval is performed by DNV GL It is found to ...

Страница 4: ...ns which could lead to possible damage to material or environmental damage NO DOMESTIC WASTE The device must not be disposed together with domestic waste IMPORTANT Use the labels on the front and back to check if the delivered unit corresponds to your order Delivery unpacking and scope of supply All units have been temperature cycled calibrated and carefully checked for their operational reliabili...

Страница 5: ... thermostats can be tested and calibrated Product identification Please refer to the label in front for mains supply and fuse Do not apply improper voltage Please refer to label at the back for type range serial number DNV GL type approval calibration information and safety instructions Relative to ambient For detailed technical information see page 16 Model Range Stability TC15 40 C to 150 C 0 05...

Страница 6: ...2 Color touch display 3 USB service interface 4 Temperature well 5 Ventilation for airflow 6 Switch test input 7 AC power 8 Fuse 9 On Off switch 10 Fan 11 Ventilation for airflow Product description 1 2 3 4 5 7 11 10 9 8 6 ...

Страница 7: ...6 5mm Ø 3 5mm 1 4 1 8 Ø 6 5mm Ø 4 5mm Ø 6 5mm Ø 6 5mm 1 4 1 4 Ø 3mm Ø 10mm Ø 4 5mm Ø 13 5mm Ø 5mm Ø 13 5mm Ø 5 5mm Ø 13mm Ø 6 5mm Ø 14 5mm Ø 7mm Ø 10mm Ø 7mm Ø 12mm Ø 8mm Ø 12mm 3 x Ø 6 5mm 3 x 1 4 6 x Ø 6 5mm 6 x 1 4 Standard inserts TC15 Ø 6 5 mm 1 4 Ø 3 5 mm 1 8 Optional inserts TC15 Undrilled Ø 4 5mm Ø 5mm 3 16 Ø 8mm 5 16 Ø 9 5mm Ø 10mm Ø 10 5mm 3 8 Ø 11mm Ø 12mm Ø 15 5mm Ø 6 5mm Ø 3 5mm 1 4 1...

Страница 8: ... down buttons multiple times in order to set the desired calibration temperature set point Press and hold one of the buttons to count up or down If the buttons are held continuously the counting speed will increase The set point is active immediately after button is released 8 When both the calibrator and calibration object is stable note both readings 9 Proceed to the next calibration point with ...

Страница 9: ...l be exposed Always ensure that the air inlet in the bottom of the TC65 TC15 is free and not clogged by dust Ensure that the fan is running If not do not use the TC65 TC15 as it will over heat and create a potential danger Return the TC65 TC15 for repair Ensure that the TC65 TC15 and especially the well is clean and free for dust before use Do not use oil in the well to get better thermal contact ...

Страница 10: ...ibration process should now ensure that your temperature instrument is within these limits and not necessarily the peak performance of the specifications of the instrument A good rule of thumb is to double the accuracy demand for your calibration compared to the process accuracy demand This leaves the accuracy demand for your calibration in the range of 2 5 C The whole chain of elements involved i...

Страница 11: ...tion you can trace back and identify the time span where errors may have occurred and take actions thereafter The shorter the calibration interval the shorter time span with possible errors A normal interval for calibrating a thermometer in a process should be about 1 3 months If it s important to ensure the consistency of the accuracy make simplified calibrations between scheduled calibrations wi...

Страница 12: ...ng 6 2 Place the probe in an appropriate insert in the well Use a hole in insert that is 0 2 to 0 6mm larger than the probe you are calibrating Insert adapters are available with a large number of dimensions It is also important that the probe reaches the bottom of the well If not the reading will be very inaccurate 6 3 Turn on the calibrator Self explanatory 6 4 Choose unit C or F The default set...

Страница 13: ...l well temperature There will also be a large difference in the actual well temperature and the temperature displayed on the calibration object UUT Therefore the temperature has to equalize After the built in stabilization criteria is reached there is a 5 minutes delay before the stable state is shown in the display Additional time must be added if the reading of the UUT is not stable This is norm...

Страница 14: ...brator inputs Please pay attention to the fact that thermostats react slowly to temperature changes Switch open or Switch closed will be shown in the display 6 11 Switch off and storage NEVER REMOVE HOT PROBES OR INSERTS FROM THE WELL Cool the calibrator to 100 C before switch off and below 50 C before removing probe and insert for storage ...

Страница 15: ...may be wise to keep a closer eye at the calibrator and maybe calibrate it more often or have it examined This also applies to your calibration objects TC65 TC15 is a dry well calibrator Do not use oil or other liquids in the well or inserts There are no user serviceable parts inside the calibrator However there is a main fuse next to the power switch that may breake due to internal failure A quali...

Страница 16: ...mm Approx 5 5 kgs TC15 0 4 C 50 min 20 min 150 C to 40 C Yes Yes 110 mm 19 mm 120 240 VAC Fuse 3AT 50 60 Hz 180 Watt 0 to 40 C 122 x 345 x 240 mm Approx 5 5 kgs Relative to ambient Relative to ambient Model Range Stability Accuracy Accuracy 1 year 3 year TC15 40 C to 150 C 0 05 C 0 3 C 1 2 C TC65 30 C to 650 C 0 1 C 1 5 C 3 5 C The calibrators have touch display and a calibration interval up to 3 ...

Страница 17: ...tten instruction describing how to maintain the calibrator In the column Standard used note the serial number for your TC65 TC15 unit Also note the number at the certificate for the latest calibration of your TC65 TC15 calibrator and the name of the company or organization that performed the calibration This way you will maintain the traceability all the way from your calibration object to the int...

Страница 18: ...libration due date Calibration conditions Temperature RH Standard used Instrument Model Serial No Cert No Traceability Temp calibrator TC65 TC15 Measuring report TC65 TC15 reading TC65 TC15 correction TC65 TC15 real value Test object reading Test object deviation Remarks Test object deviation is calculated by subtracting TC65 TC15 real value from Test object reading The value Test object deviation...

Страница 19: ...TC65 TC15 PS00526 Version 2 2020 Nemko Test Report Electromagnetic Compatibility Radiated RF Disturbance Immunity EN 61326 1 2013 DNVGL CG 0339 2019 EN 61000 4 3 2010 Ed 3 2 Observations show that the temperature control is disturbed outside the specifications when subject to a field strength of 10V m in the frequency range 80 1000 MHz The performance returns to normal when the field is switched o...

Страница 20: ...IKM Instrutek AS Main office Elveveien 28 3262 Larvik Norway T 47 33 16 57 00 F 47 33 16 57 01 E Instrumentsalg IKM no www IKM com ikm instrutek ...