Maintenance and cleaning

The device is maintenance-free. However it is subject to natural 
wear and tear on parts and their statistical failure rate.

Unplug from the mains before cleaning.

Only clean the 

IKA

 device using these 

IKA

  approved cleaning 

agents:

Dirt

Cleaning agent

Dyes

Isopropyl alcohol

Building materials

Water containing detergent, isopropanol

Cosmetics

Water containing detergent, isopropanol

Food

Water containing detergent

Fuels

Water containing detergent

Other materials

Please ask 

IKA

• 

Wear protective gloves when cleaning the devices. 

• 

Do not place electrical appliances into the cleaning agents for 
cleaning purposes.

• 

Ensure no liquid enters the device during cleaning.

• 

Please consult with 

IKA

 before using any cleaning or decon-

tamination methods not specifically recommended.

Ordering spare parts
When ordering spare parts, please make sure to indicate the fol-
lowing:
• 

Device type

• 

Device manufacturing number; see rating plate

• 

Reference number and description of spare part; see spare 
part diagram and list at 

www.ika.de

.

Repairs

Please only send devices in for repair that have been 

cleaned and are free of materials which might present 

health hazards.

For this, use the 

“certificate of compliance”

 form which you 

can obtain from 

IKA 

or can download a version for printing from 

the 

IKA

 website 

www.ika.de

If your appliance requires repair, return it in its original packaging. 
Storage packaging is not sufficient. also use appropriate transport 
packaging.

1
2 RxD

3 TxD

4

5 GND

6

7 RTS

8 CTS

9

1

RxD 2

TxD 3

4

GND 5

6

RTS 7

CTS 8

9

PC

1  

2  

3  

4  

5

6  

8  

9

9  

8  

7  

6

5  

4  

3  

2  

1

“Watchdog” functions; monitoring of the serial data flow

If, once this function has been activated (see NAMUR commands), 
there is no retransmission of the command from the computer 
within the set time (“watchdog time”), the Evacuation function is 
switched off in accordance with the set “watchdog” mode or is 
returned to previously set target values. The data transmission may 
be interrupted by, for example, a crash in the operating system, 
a power failure in the PC or an issue with the connection cable 
between the computer and the device.

“Watchdog” mode 1
If event WD1 should occur, the evacuation function is switched 
off and WD1 Watchdog Error is displayed. Set watchdog time to 

m (20 - 1,500) seconds, with watchdog time echo. This command 
launches the watchdog function and must be transmitted within 
the set watchdog time.

“Watchdog” mode 2
If there is an interruption in data communications (longer than the 
set watchdog time), the speed target value is changed to the set 
WD safety speed limit. The warning PC 2 is displayed. The WD2 
event can be reset with OUT_WD2@0 - this also stops the watch-
dog function.
Set watchdog time to m (20 - 1,500) seconds, with watchdog time 
echo. This command launches the watchdog function and must 
be transmitted within the set watchdog time.

Содержание Vacstar control

Страница 1: ...20000023248 IKA Vacstar control IKA Vacstar control_102018 Operating instructions EN ...

Страница 2: ...an Ventilation slit 7 Hose connection for suction line d 8 mm INLET 8 USB interface 9 RS 232 interface 10 Water Valve Connection 11 Hose connection for pressure line d 8 mm OUTLET 12 Mains cable connection 13 Pressure Device 14 Hose connection for ventilation line d 8mm 15 Hose connection for load line d 8mm 16 Hose connection for suction line d 8 mm INLET 17 Clamp 18 Receiving flask 1 2 3 4 14 12...

Страница 3: ...9 1 EN 301489 17 EU Declaration of conformity Explanation of warning symbols According to IKA s Terms and Conditions of sale and delivery this product is covered by a warranty for a period of 24 months Please contact your dealer for any warranty claims If you wish you can send the device directly to our factory Please provide the sales invoice and state the reasons for your guarantee claim In this...

Страница 4: ...ore commissioning check that all the housing parts are present and fastened to the device Do not lift the pump if the handle is loose and the handle securing screw Fig 2 5 is loose or missing Only switch ON the pump if the pump is standing vertically Connect the hose connections INLET OUTLET and interfaces in accordance with the labelling on the device and the operating instructions Ensure that th...

Страница 5: ...thium polymer battery pack must only be used and charged in IKA products designed for use with this battery pack When the battery pack is inserted it should slide in easily and without resistance Do not force it If the battery pack is removed for an extended period of time store it in a sealed plastic bag to prevent short circuiting due to moisture or contact with metal components The operating te...

Страница 6: ...Vacstar control Power cord set USB A USB B cable A USB A USB micro B Vacuum hose 1 m Vacuum Hose 0 3 m connection Vacstar to Vacuum Safety VSE 1 Vacuum Safety VSE 1 Wireless Controller WiCo OS 1 0 power supply unit for WiCo Adapter UK Adapter USA China Adapter Australia Adapter Europe Switzerland Cover Screw driver screw driver to screw WiCo on station Bracket Flask Spherical clamp User guide Warr...

Страница 7: ...h no load at the end of the process no load operation approx 3 up to max 5 minutes to dispel any residues in the pump If applicable vent the recipient To do this remove the suction line during manual operation NOTE Observe the operating instructions of the relevant devices Area of use indoor use only Laboratories Schools Pharmacies Universities The device is suitable for use in all areas EMC class...

Страница 8: ...the station the rechargeable battery is automatically recharged via the contacts page 11 A Q The rechargeable battery can also be charged via the USB port of the WiCo In automatic operation with controlled speed the pump reduces its speed when it reaches the target value If the leakage rate in the entire system is low enough the speed may even be set to zero Schematic view of speed vacuum control ...

Страница 9: ...elling on the pump and the operating instruc tions The connections are barbed hose connectors Ensure that the hos es have hose clips if applicable Lay the hoses in such a way that no condensate can flow into the pump Ensure that the air can enter the ventilation slit Fig 2 6 unhin dered Install a condensate separator e g a VSW 1 in the suction line before the intake spigot of the pump to protect a...

Страница 10: ...bsorber to the end NOTE Put the end of the hose in the fume hood Check there is a free outlet on the pressurised side Do not use a throttle on the pres surised side and do not close the outlet Connect the exhaust line to this connection Fig 2 8 USB interface Connect the vacuum pump IKA Vacstar control to a PC with a USB A USB B cable Any device software updates can be loaded using the IKA FUT soft...

Страница 11: ... the settings in the menu L Securing screw Secure the Wico to the station M On Off key Switching the Wireless Controller on and off N Lock key Key disable and rotary push knob disable O Vent key To vent the load if necessary during the vacuum process P Pause key Temporarily stop the function of the device for required period Note can resume the function by pressing the button again Q PCB Contact W...

Страница 12: ...ions or optimisations of previous functions You can download the Firmware Update Tool at www ika de fut Working screen factory settings The following working screen appears automatically in the display Explanation of symbols on the working screen Lock key This symbol means that the function of the key and of the rotary knob for controlling the vacuum controller are disabled The sym bol no longer a...

Страница 13: ...urn to the previous menu or working screen Note The menu option activated is highlighted blue in the display Menus which are grayed out are inactive Navigation menu Press the Menu key H and turn the rotary push knob K Press the Back key I or the Menu key H Note If you press the Menu key H the system skips directly back to the working screen If you press the Back key I the system skips back to the ...

Страница 14: ... update info Communication Brightness Pressure max Operating mode Logic version Volume Pressure min Pump max Pump min Key tone Device name Firmware Update Info Display version Activated Deactivated 20 05 00 mm ss Add solvente No 1 5 Pump Display Black White Background Standard Mode Battery Mode Sound Factory setting Menu values Programs All Bluetooth Remote control Paring Information Deactivated D...

Страница 15: ...g the knob again stops the process Cleaning NOTE The Cleaning menu item increases the service life of the pump since solvent residues do not remain in side the pump The cleaning of the pump ensures that solvent residues are re moved in the pump after the measurement For this purpose the pump should run at a low speed rate while the system is venting The air flowing through the pump ensures that th...

Страница 16: ...tion for a standard time The device can emit a signal beep after the set target time has elapsed Operating mode 1 Press the Menu key N 2 Select the menu Operating mode and confirm the input 3 Select the following settings Operating mode A In this operating mode the set target value is not saved when the current run comes to an end or the device is switched off Limits In this menu limits for the ta...

Страница 17: ... segment in the program Delete Deletes the selected program If a selected program is deleted by pressing on menu option Delete with rotary push knob all the program parameters will be emptied The tick disappears Details for editing the program In this program you can define up to 10 segments The selected segment is highlighted Then you can edit delete or insert a seg ment in this program The progr...

Страница 18: ...t In the Time Out menu you can set a time limit This time limit applies if there is a communication failure between the IKA Vacstar and the WiCo or the communication range is exceeded The IKA Vacstar will continue to run at the set speed until the set time has expired After that the IKA Vacstar will continue to run at the set safety pressure manual mode or safety pump rate Pump mode see the Safety...

Страница 19: ... to choose menu values pro grams and all Select the option which is going to be reset Communication The Device name option allows you to input a device name The device name appears on the opening screen when the device is switched on This can be useful when using multiple devices with different set tings This also facilitates the identification of each device when communicating with a PC via USB R...

Страница 20: ... space code hex 0x20 Each individual command including parameters and data and all responses are completed with CRLF code hex 0x20 hex 0x0d hex 0x0A and can have a maximum length of 50 characters The dot is used for decimal separators in a floating point val ue code hex 0x2E The details given above generally comply with the recommen dations of NAMUR NAMUR recommendations for the design of electric...

Страница 21: ...h you can obtain from IKA or can download a version for printing from the IKA website www ika de If your appliance requires repair return it in its original packaging Storage packaging is not sufficient also use appropriate transport packaging 1 2 RxD 3TxD 4 5 GND 6 7 RTS 8 CTS 9 1 RxD 2 TxD 3 4 GND 5 6 RTS 7 CTS 8 9 PC 1 2 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 Watchdog functions monitoring of the seria...

Страница 22: ...l service department Error 8 Speed sensor fault Device can t detect the pump speed The sensor occurs some unknown errors that device can t read the speed value Call service department Error 9 Storage Error Read or write internal flash error Read or write internal flash error Call service department In the event that the fault cannot be eliminated using the measures described or another error code ...

Страница 23: ...um sensor type ceramic Al2O3 Pressure max for pressure sensor bar 1 2 Measurement range absolute min mbar 1 Measurement range absolute max mbar 1030 Resolution pressure mbar 1 Measurement uncertainty mbar 1 Venting valve yes Condensate trap yes Emission condenser yes Timer yes Noise at low pressure dB A 54 Material in contact with medium Al2O3 PTFE FPM PPS Housing material Alu cast coating Mode au...

Страница 24: ...ax C 40 Permissible relative humidity 80 Protection class according to DIN EN 60529 IP 20 USB interface yes RS 232 interface yes Voltage V 100 240 Frequency Hz 50 60 Power input W 140 Power input standby W 1 5 Average values Subject to technical changes ...

Страница 25: ...25 ...

Страница 26: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com ...

Отзывы: