background image

26

Error codes

Any malfunctions during operation will be identified by an error message on the display�
Proceed as follows in such cases:

 Switch off device using the main switch at the back of the device�

 Carry out corrective measures�

 Restart device�

Error code

Effect

Cause

Solution

Error 01

Pump off
Refrigerating off

No external temperature sensor

- Check this sensor

Error 02

Pump off
Refrigerating off

Motor over current (rate current) - Reduce pump motor speed

- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Error 04

Pump off
Refrigerating off

Motor hall signal missing

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Error 06

Pump off
Refrigerating off

Too low liquid level

- Check the liquid level and buoyage

Error 10

Pump off
Refrigerating off

PC communication failure 

- Check communication cable

Error 15

Pump off
Refrigerating off

Compressor power supply 
disconnected

- Restart the device

Error 16

Pump off
Refrigerating off

Suction temperature error 

 

- Restart the device

Error 17

Pump off
Refrigerating off

Discharge temperature  error  

- Restart the device

Error 19

Pump off
Refrigerating off

Fluid temperature too high

- Switch the device off and let the fluid cool down

Error 20

Pump off
Refrigerating off

Condenser fan error

- Restart the device

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps: 
- Contact the service department
- Send the device for repair, including a short description of the fault� 

Содержание RC 2 GREEN control

Страница 1: ...Operating instructions EN IKA RC 2 GREEN control...

Страница 2: ...ig 1 Item Designation 1 Filling opening lid 2 Wireless Controller WiCo 3 Handle 4 Venting grid 5 RS 232 port 6 USB port 7 External temperature sensor socket 8 Pump connector IN 9 Pump connector OUT 10...

Страница 3: ...g battery into the WiCo 9 Mounting the WiCo to the station 9 Filling and draining 9 Fluids Standard information for IKA fluid 11 Moving the device 12 Charging the Battery Pack RB 1 rechargeable batter...

Страница 4: ...nse exempt RSS standard s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device must accept any interference including interference that may cause...

Страница 5: ...on the device again to ensure safe operation for example to keep foreign ob jects and liquids etc from getting into the device The device may only be used as prescribed and as de scribed in these ope...

Страница 6: ...OFF key and the Power switch Please note the following safety instructions for the battery pack RB 1 recharge able battery Keep the battery pack out of reach of children at all times Store the batter...

Страница 7: ...k between the WiCo and the laboratory device first check whether your region is in cluded in the radio communication approval for the de vice If it is not remote control can also be performed using a...

Страница 8: ...he cooling system Connecting the tubings Unscrew the union nuts and stoppers using a wrench AF 19 from the pump connector IN 8 and OUT 9 Connect the hoses for circulating the external system to the pu...

Страница 9: ...w cap Inserting battery into the WiCo Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Note notch upward Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Fig 7 Scr...

Страница 10: ...ckwise to the stop position see Fig 10 Note Please note information in section Commissioning Connect the power plug and turn on the device with power switch 10 The low level icon appears on the displa...

Страница 11: ...perature Maximum temperature N Negative Temperature 4 Viscosiy Viscosiy at 25 C for Heating Fluid HF Viscosity at 20 C for Cooling Fluid CF Viscosity at 25 C for Universal Fluid UF Dynamic viscosity m...

Страница 12: ...ss head clamp H 38 Holding rod IKA RC 2 GREENcontrol IKA RC 2 GREEN control IKA RC 2 GREENcontrol Moving the device Empty all fluid in the bath before move device from one place to other place The dev...

Страница 13: ...gh Low Fluid level 2 GREEN Fig 16 A F D E B C Item Designation Function L On Off button Switch the WiCo on and off M int ext button Switch between the internal and external temperature display and con...

Страница 14: ...nce The refrigerating function may delay to start at maximum 2 minutes In working status press the right knob R to stop opera tion of the pump The refrigerating function and the pump stops Note In sta...

Страница 15: ...and the station The screen below shows the most significant symbols on the working screen Bluetooth This symbol means the station and the WiCo are com municating via Bluetooth The symbol no longer app...

Страница 16: ...he menu by turning the right rotating pressing knob R to the right or left Open the menu item by pressing the right rotating pressing knob R Turn the rotating pressing knob R to select the desired men...

Страница 17: ...omized 1 Customized 2 PUMP Limits Speed Pressure Maximum Maximum Minimum 4000 rpm 2000 rpm 0 5 bar TIMER Set Display 00 00 hh mm MODE A C B D activated GRAPH Automatic Manual Axis Assignment Set Tempe...

Страница 18: ...e difference between the external system and internal bath during the heating up and cooling down process This can protects sensitive device like the glass reactors from thermal shock e g glass break...

Страница 19: ...Step 1 2 Enter the temperature from the measuring device 32 00 OC OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the measuring device 32 00 OC OK Calibration Internal...

Страница 20: ...d or set PROGRAMS Under programs 10 user defined temperature time pro files can be created A program can consist of up to 10 segments Once a program has been selected the following options are availab...

Страница 21: ...button re sets all the system settings to the original standard values set at dispatch from the factory see Menu structure 5 Communication The Device name menu option allows you to edit the device na...

Страница 22: ...top mode Type of transmission full duplex Character format character representation in accordance with data format in DIN 66 022 for start stop mode 1 start bit 7 character bits 1 parity bit even 1 st...

Страница 23: ...1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the cooling and pump functions are switched off and message PC 1 is display This command needs to...

Страница 24: ...o A USB micro B USB micro A Connection the device to PC USB micro B USB A USB micro B USB A RS 232 USB or USB micro B USB A Fig 32 USB cable RS 232 USB or USB micro B USB A 9 pin RS 232 9 pin RS 232 A...

Страница 25: ...alcohol Wear protective gloves while cleaning the instruments Electrical instruments may not be placed in the cleansing agent for the purpose of cleaning Do not allow moisture to get into the instrume...

Страница 26: ...impeller is blocked Error 06 Pump off Refrigerating off Too low liquid level Check the liquid level and buoyage Error 10 Pump off Refrigerating off PC communication failure Check communication cable E...

Страница 27: ...h 12 H SI 8 Silicone hose nominal width 8 H SI 12 Silicone hose nominal width 12 H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12...

Страница 28: ...Knob on WiCo Temperature setting resolution K 0 1 Temperature display TFT LCD on WiCo Temperature display resolution K 0 01 Cooling capacity according to DIN 12876 at 4000 rpm 20 C 10 C 0 C 10 C 20 C...

Страница 29: ...ranty conditions the warranty period is 24 months For claims under the warranty please contact your local dealer You may also send the machine direct to our factory enclosing the delivery invoice and...

Страница 30: ...es england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22...

Отзывы: