20
MODE
1� Operating Mode A:
After power-on/power failure no automatic restart of
functions�
2� Operating Mode B:
After power-on/power failure automatic restart of func-
tions, depending on previous settings�
3� Operating mode C:
Set values (set in A or B) cannot be changed�
After power-on/power failure automatic restart of func-
tions, depending on previous settings�
4� Operating mode D:
Confirmation request for set value changes, if functions
are active� After power-on/power failure no automatic re-
start of functions�
DISPLAY
1� Main Screen:
In the “Main Screen“ menu option, you can specify what
information will be displayed on the screen�
2� Device Number:
In the “Device number“ menu option, you are allowed to
edit the device number from 0 to 99 that will be shown
on the main screen�
When using multiple units, it may be helpful to identify
WiCo and station�
Note:
Labeling device number to the station�
GRAPH
In this menu, you can set the options for the time-temper-
ature diagram�
1� Automatic:
The scaling of the temperature axis (Y-axis) is automatically
determined, depending on the target temperature and the
actual internal and external temperature�
The time axis (X-axis) is permanently scaled to 30 minutes�
2� Manual:
1) Axis Assignment:
The temperature values to be displayed can be selected�
2) Axis scaling:
Scaling of the time (X) and temperature axis (Y) can be
selected or set�
PROGRAMS
Under programs, 10 user-defined temperature-time pro-
files can be created� A program can consist of up to 10
segments�
Once a program has been selected, the following options
are available:
1� Start:
Starts the program upon request of loop mode�
1) Infinite loop:
Upon completion of the last segment,
the program continues with the first segment until the
user ends the program by stopping a device function�
2) Loop Count:
Indicates the total number of loop cycles
until program end�
Note:
At the end of the program all device functions are
switched off�
2� Edit:
Edit/change the selected program�
1) Seg No.:
Segment number�
2) Ctrl. Sensor (int/ext):
Determines whether control is
through the internal (int) or external (ext) temperature
sensor�
3) Temp.:
Target temperature�
4) Ctrl.Mode (Time / +/- x.x K):
In “
Ctrl�Mode Time
”
the target values and settings of the segment are valid
for the duration indicated in the column “
Time hh:mm
”�
Afterwards, the next program segment is automatically
executed�
In “
Ctrl�Mode +/- x�xx K
”, the hysteresis (tolerance) of
the actual temperature to the target temperature is set
(e�g� +/- 0�1 K)� The target values and settings of the seg-
ment are valid until the actual temperature reaches the
target tempe/- hysteresis for the first time�
Afterwards, the next program segment is automatically
executed�
5) Pump rpm:
Target speed of pump�
Status” as “Open”
Edit:
Edit/change the selected program parameters�
Delete:
Delete the selection highlighted in yellow pro-
gram segment�
Insert:
Inserts a new program segment after the selected
segment�
Save:
Saves changes�