RO105STS-0310E
1-1, Ishida, Toyooka-cho, Gamagori-shi, Aichi 443-0011 JAPAN
https://www.iijima-e.co.jp/
1. Calibrating the span
5. Replacing the needle
6. Replacing the membrane filter
7. Replacing the check valve
7. Replacing the oxygen sensor
1. Release the needle to the atmosphere.
The Menu screen opens.
2. Press .
3. Press .
4. Hold down for 1 second or longer.
At this time, confirm that the O-ring for housing is
properly mounted on the groove.
The calibration starts, and the Stability Decision screen opens.
A beep sounds, “CAL complete” appears (approx. 1 second), and then the O
2
MEASURE
screen opens. The calibration process is completed.
The calibration value is saved even if the power is turned OFF, the batteries are removed,
or AC adaptor is disconnected.
If the “Sensor Error” message appears, refer to the section “Error Messages” in the operation
manual enclosed with the product, and confirm and remedy the situation.
-
Once a day, calibrate the span before taking the first measurement. (There is no need to calibrate before each
measurement.)
- Ideally, calibrate when the oxygen sensor temperature and temperature sensor temperature are sufficiently
stable, such as first thing in the morning, indoors before turning on the air-conditioning.
Residual Oxygen Meter
PACK KEEPER
Model:RO-105KS
PACK MASTER
®
Model:RO-105S
PACK LEADER
Model:RO-105LS
Handy
User’s Guide
Residual Oxygen
/
Carbon Dioxide Meter
<Note>
This is a handy user’s
guide. Please refer to the instruction
manual enclosed with the product
for more details on the operation
and precautions.
<Handling and Maintenance Information>
Information and videos on this product’s
handling methods and maintenance, etc.,
are available on the IIJIMA ELECTRONICS
website.
Progress bar
* Order the needle (Model:NN-2116R-12 / vertical
hole type, 12-pc. set)
* Order the menbrane filter
(RO-MF,5-pc. set)
* Order the check valve
(RO-CV, 3-pc. set)
* Order the oxygen sensor
(WAGNIT
®
Model:WA-SGF).
Check valve
Sampling tube
with fitting
Main cover
Oxygen
sensor
Housing
O-ring for
housing
Probe
Hold the rotating
section with fingers
so that it does not
spin while removing
the membrane filter.
Membrane filter
(1) Remove the needle from the membrane filter.
(1) Remove the needle from the membrane filter.
(2) Remove the membrane filter from the check valve.
(3) Attach the new membrane filter.
(4) Attach the needle.
(1) Remove the membrane filter and needle from the
check valve.
(1) Remove the main cover.
(3) Pull the sensor out of the probe.
(4) Align the new sensor with the insertion hole
on the probe, and insert it.
(6) Mount the main cover.
(5) Turn the sensor clockwise and align it to the
mark on the housing.
(2) Turn the sensor counterclockwise and
pull it straight out.
(2) Remove the check valve from the samplig tube
with fitting.
(3) Attach the new check valve onto the samplig tube
with fitting.
(4) Attach the membrane filter and needle.
(2) Attach a new needle.
Dispose of the spent needle according to local rules.It cannot be reused.