background image

RO105STS-0310E

1-1, Ishida, Toyooka-cho, Gamagori-shi, Aichi 443-0011 JAPAN 

https://www.iijima-e.co.jp/

1. Calibrating the span

5. Replacing the needle

6. Replacing the membrane filter

7. Replacing the check valve

7. Replacing the oxygen sensor

1. Release the needle to the atmosphere.

The Menu screen opens.

2. Press                  .

3. Press                  .

4. Hold down                  for 1 second or longer.

At this time, confirm that the O-ring for housing is 

properly mounted on the groove.

The calibration starts, and the Stability Decision screen opens.

A beep sounds, “CAL complete” appears (approx. 1 second), and then the O

MEASURE 

screen opens. The calibration process is completed.

The calibration value is saved even if the power is turned OFF, the batteries are removed,

or AC adaptor is disconnected.

If the “Sensor Error” message appears, refer to the section “Error Messages” in the operation

manual enclosed with the product, and confirm and remedy the situation.

Once a day, calibrate the span before taking the first measurement. (There is no need to calibrate before each

   measurement.)

- Ideally, calibrate when the oxygen sensor temperature and temperature sensor temperature are sufficiently

  stable, such as first thing in the morning, indoors before turning on the air-conditioning. 

Residual Oxygen Meter

PACK KEEPER

Model:RO-105KS

PACK MASTER

®

Model:RO-105S

PACK LEADER

Model:RO-105LS

Handy

User’s Guide

Residual Oxygen

/

Carbon Dioxide Meter

<Note>  

This is a handy user’s

guide. Please refer to the instruction 

manual enclosed with the product 

for more details on the operation 

and precautions.

<Handling and Maintenance Information> 

Information and videos on this product’s

handling methods and maintenance, etc., 

are available on the IIJIMA ELECTRONICS 

website.

Progress bar

* Order the needle (Model:NN-2116R-12 / vertical

   hole type, 12-pc. set)

* Order the menbrane filter 

   (RO-MF,5-pc. set)

* Order the check valve

   (RO-CV, 3-pc. set)

* Order the oxygen sensor 

   (WAGNIT

®  

Model:WA-SGF).

Check valve

Sampling tube 

with fitting

Main cover

Oxygen

sensor

Housing

O-ring for

housing

Probe

Hold the rotating 

section with fingers 

so that it does not 

spin while removing 

the membrane filter.

Membrane filter

(1) Remove the needle from the membrane filter.

(1) Remove the needle from the membrane filter.
(2) Remove the membrane filter from the check valve.
(3) Attach the new membrane filter.
(4) Attach the needle.

(1) Remove the membrane filter and needle from the 

     check valve.

(1) Remove the main cover.

(3) Pull the sensor out of the probe.

(4) Align the new sensor with the insertion hole

     on the probe, and insert it.

(6) Mount the main cover.

(5) Turn the sensor clockwise and align it to the

     

   

  mark on the housing.

(2) Turn the sensor counterclockwise and

     pull it straight out.

(2) Remove the check valve from the samplig tube 

     with fitting.
(3) Attach the new check valve onto the samplig tube 

     with fitting.
(4) Attach the membrane filter and needle.

(2) Attach a new needle.

Dispose of  the spent needle according to local rules.It cannot be reused.