background image

 

17 

 

Table 3 – Proximity fields from RF wireless communications 

equipment

 

 

Test frequency 

 

(MHz)

 

Band

 

(MHz)

 

Immunity test levels

 

Professional healthcare facility environment

 

385

 

380-390

 

Pulse modulation 18Hz, 27V/m

 

450

 

430-470

 

FM, 

±

5kHz deviation, 1kHz sine, 28V/m

 

710

 

704-787

 

Pulse modulation 217Hz, 9V/m

 

745

 

780

 

810

 

800-960

 

Pulse modulation 18Hz, 28V/m

 

870

 

930

 

1720

 

1700-1990

 

Pulse modulation 217Hz, 28V/m

 

1845

 

1970

 

2450

 

2400-2570

 

Pulse modulation 217Hz, 28V/m

 

5240

 

5100-5800

 

Pulse modulation 217Hz, 9V/m

 

5500

 

5785

 

WARRANTY 
Please contact your dealer or the device center in case of a claim under the 
warranty. If you have to send in the unit, enclose a copy of your receipt with 
clear statement of defect description. 
The warranty terms as below: 
1. The warranty period for device is one year from date of delivery. In case of a 
warranty claim, the date of delivery has to be proven by means of the sales 
receipt or invoice. 
2. Repairs under warranty do not extend the warranty period. 
3. The following cases are excluded under the warranty 

• 

All damage  which has arisen due to improper  treatment, e.g. 
nonobservance of the user instruction. 

• 

All damage which is due to repairs or tampering by the  customer or 
unauthorized third parties. 

• 

Damage which has arisen during transport from the  manufacturer to the 
consumer or during transport to the service center. 

• 

Accessories which are subject to normal wear and tear. 

4. Liability for direct or indirect consequential losses  caused by the unit is 
excluded even if the damage to the unit is accepted as a warranty claim. 

 

Manufactured for iHealth Labs Inc. 
120 San Lucar Ct., Sunnyvale, CA 94086, USA 
+1-855-816-7705 

Содержание PT3SBT

Страница 1: ...1 iHealth PT3SBT Wireless No Touch Forehead Thermometer FCC ID SLRPT3SBT IC 10913A PT3SBT...

Страница 2: ...applications 8 Contraindications 8 Instruction For Use 8 SYNC TIME AND RESULTS VIA BLUETOOTH 11 Product errors and troubleshooting 12 Body temperature 12 Cleaning and Care 13 Care the probe 13 Mainten...

Страница 3: ...bout the temperature reading Keep the thermometer out of reach of children For accidental swallowing of the battery or other components please contact emergency services immediately Batteries must not...

Страница 4: ...y cause errors in the reading or even device failure 12 Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 i...

Страница 5: ...owing occurred 1 Operation outside of the manufacturer specified subject temperature range 2 Operation outside of the manufacturer specified operating temperature and humidity ranges 3 Storage outside...

Страница 6: ...ncluding interference that may cause undesired operation of the device Under Industry Canada regulations this radio transmitter may only operate using an antenna of a type and maximum or lesser gain a...

Страница 7: ...s 4 Measurement range 89 6 109 2 32 42 9 5 Measurement precision 0 4 0 2 within 93 2 107 6 34 42 and 0 5 0 3 for other temperature ranges 6 Resolution 0 1 0 1 7 Clinical reproducibility Within 0 5 0 3...

Страница 8: ...ntraindications It is not recommended for people whose measuring part has local lesions such as inflammation trauma postoperative etc Instruction For Use The patient is an intended operator The patien...

Страница 9: ...ead If the icon on the thermometer screen is always on the measured temperature data can be uploaded to the smart device Possible reasons for measurement errors include A Measurement distance is great...

Страница 10: ...if the battery capacity is too low for measurements the screen will display a single flashing icon and automatically switch OFF after 8 seconds To continue using the device old batteries must be repla...

Страница 11: ...ress the button to turn on the thermometer or start one measurement 2 Enable Bluetooth on your Android device 3 When using the thermometer for the first time you should pair the thermometer to the And...

Страница 12: ...g Target temperature is beyond range of measurements 89 6 109 2 32 42 9 Environment temperature either exceeds the design range 32 104 0 40 or is unstable Follow the instruction manual and repeat the...

Страница 13: ...ermometer casing thoroughly The wiping process preferably lasts 15 seconds Allow at least 5 minutes drying time before taking a temperature After cleaning if the device is not visually clean when obse...

Страница 14: ...E Directive If you have any queries please refer to the local authorities responsible for waste disposal Notes Please act according to local laws for disposal of used batteries Take out the battery if...

Страница 15: ...h is a trademark of iHealth Labs Inc The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc And any use of such marks by iHealth Labs Inc is under license Calibration T...

Страница 16: ...n tested against this SAR limit ELECTROMAGNETIC COMPATIBILITY INFORMATION Frequency range Bluetooth 2402MHz 2480MHz Max EIRP 4 40dBm Table 1 Emission limits per environment Phenomenon Compliance Elect...

Страница 17: ...ty terms as below 1 The warranty period for device is one year from date of delivery In case of a warranty claim the date of delivery has to be proven by means of the sales receipt or invoice 2 Repair...

Страница 18: ...ict Tianjin 300190 China Tel 86 22 60526161 This product can be used across EU member states EU Regulatory Conformance Hereby iHealth Labs Inc declares that this device is in compliance with the essen...