STERYLIS ULTRA Operating Manual
Replacing the filters
Pre-filter replacement (F1)
Active carbon filter replacement (F2)
Replacing the UV-C lamps
Regular filter replacement is necessary for proper operation. This applies to both pre-filters and activated
carbon filters. The condition of the filters should be checked every 500 operating hours. The filters must
be replaced every year (unless the filters show no signs of use). The need to check/replace the filters is
indicated by appropriate messages. In addition, it is necessary to replace the activated carbon filter with a
standard filter when the device is in sterilisation mode.
Make sure that the device is switched off and disconnected from the power supply.
Remove the wind girder, lift the filter gently up, bend it slightly, pull it down. Install the new filter in
the same way, but in reverse order.
A pre-filter that is pulled out of the destination causes a violation of the limiting switch. If a limiting
switch is violated during the sterilisation process, the process will be stopped and completed and the
device will go into standby mode. In case of violation of the limiting switch in filtration or disinfection
mode, the process will be stopped at the current moment and the unit will go into pause mode. When
the limiting switches of the filter cartridge are no longer violated (inserting the filter cartridge back),
the unit will resume the filtration/disinfection process. Violation of the limiting switch also results in a
corresponding error on the display in the “O3 sterilisation” section. If the end point violation stops, the
error is automatically reset.
The safety limiting switches provide protection in case of an attempt to change the air filters during
operation.
Make sure that the device is switched off and disconnected from the power supply.
Pull out the cartridge from the top of the device, remove the filter, insert a new filter, insert the cartridge
into the target location.
The active carbon filter that is pulled out of the destination causes a violation of the limiting switch. If a
limiting switch is violated during the sterilisation process, the process will be stopped and completed and
the device will go into standby mode. In case of violation of the limiting switch in filtration or disinfection
mode, the process will be stopped at the current moment and the unit will go into pause mode. When
the limiting switches of the filter cartridge are no longer violated (inserting the filter cartridge back),
the unit will resume the filtration/disinfection process. Violation of the limiting switch also results in a
corresponding error on the display in the “O3 sterilisation” section. If the end point violation stops, the
error is automatically reset.
The safety limiting switches provide protection in case of an attempt to change the air filters during
operation.
Regular replacement of the UV-C radiators is essential for proper operation. Exceeding the service life of
UV light bulbs may result in their burning out. It should also be remembered that with time the maximum
dose of lamp radiation decreases.
The lifetime of UV bulbs is approx. 9000 hours. This is at the same time the time after which a message
indicating that this time has been exceeded will be displayed.