STERYLIS ULTRA Operating Manual
Parameter code
Description
Min value
Max value
Default value
C101
Operating concentration in AUTO mode [ppm]
1
10
6
C102
Ozone time in AUTO mode [min]
1
360
180
C130
Siren operation in sterilisation mode [ON/OFF]
0
1
1
Parameter code
Description
d102
UV-C lamp service interval timer [s]
d105
Fan service interval timer [s]
d108
Service interval timer for the 1. ozone generator bank [s]
d111
Service interval timer for the 2. ozone generator bank [s]
d114
Ozone sensor service interval timer [s]
d117
Ozone sensor life timer [s]
The individual parameters are changed using the „+” and „-” buttons in the “O3 sterilisation” section.
The code of the selected parameter is displayed in the „O3 sterilisation” section. The current value of the
selected parameter is displayed in the „O3 concentration” section. To change the value of the parameter,
operate the „+” and „-” buttons in the „O3 Concentration” section. The active digit is selected using the
„SET” button in the same section of the user panel. To confirm the value of the selected configuration
parameter, press and hold the „PROG.SET” button. The selected value will be confirmed by a short beep
and all displays will flash.
To return to the standby mode, press and hold the „SET” button in the „Ozone time” section.
The individual parameters are changed using the „+” and „-” buttons in the “O
3
sterilisation” section.
The code of the selected parameter is displayed in the „O
3
sterilisation” section. The current value of the
selected parameter is displayed in the „O
3
concentration” section. To reset the selected counter, press and
hold the „PROG.SET” button. The reset will be confirmed by a short beep and all displays will flash.
Working time counters
The device is equipped with the operating time counters listed in the table below. The user has the
possibility to read the indications of these counters and reset them.
To view or reset the counters, press and hold the „SET” button in the „Ozone time” section while the unit is
in standby mode. Then enter the appropriate code to change the parameters.
Entering the access code:
The 4-digit code entered is shown on the display under „O3 sterilisation”. The „+” and „-” buttons change
the value by 1 active digit. An active number is the one that flashes at a high frequency, inactive numbers
light up constantly. The „PROG.SET” button changes the active digit to the next one on the right. After
entering the correct code, confirm it by pressing the yellow button „START” longer. If the code is incorrectly
entered, the code „XXXX” or „HHHH” will be displayed and a short beep will occur. In this case, the code
must be re-entered.
You can interfere with the following counters: