12
Doc Ref: SL-065 v1.2 pg 12
Calibration
Zero
Gas
Gas
Zero and Calibration
The following diagram shows the preferred method to
introduce zero and calibration gases to the sampler.
The sampler must not be over or under pressurised
during the zero or calibration process.
With this arrangement set the flow rate from the gas
cylinder such that the flowmeter shows 0.5L/Min of gas
flow spilling off.
This arrangement allows the sampler to take its required
flow rate at close to atmospheric conditions allowing
excess gas to spill off.
Always zero first and then calibrate.
With the Android device
connected to the sampler
comm port as previously
indicated select the Zero/Cal
option