29
UK
Select parameter
1� Press [Enter] briefly
2� Press [▲] or [▼] until the requested
parameter is displayed
.
2
1
Display the parameter value
3� Press [Enter] briefly
> The currently set value is displayed for
30 s�
By pressing [Enter] briefly several
times, the display switches between
parameter and parameter value�
3
Switching to the process value display
► Wait for 30 seconds
or
► Change from the submenu to the main menu, from the main menu to the process value
display with [▲] or [▼].
12 Troubleshooting
The unit has many self-diagnostic options� It monitors itself automatically during
operation�
Warnings and error states are displayed, even when the display is switched off�
Error indications are also available via IO-Link�
Display Type
Description
Fault correction
IOE�n
Error
• Unit faulty / malfunction
► Replace the unit�
SEnS
Warning Sensor signal invalid�
• Measuring pipe not sufficiently
filled�
• Medium with a too low con-
ductivity�
► Check installation position
► Check conductivity of the
media (≥ 20 µS/cm)
Err
Error
• Unit faulty / malfunction�
• Measured value greater than
130 % MEW�
• Measured value lower than
-130 % MEW�
► Replace the unit�
► Check flow range /
temperature range�