16
8.4.2 More settings
ou2
Output configuration for analogue output (OUT2):
[I] = measuring range is provided as 4...20 mA \[InEG] = measuring
range is provided as 20...4 mA\[OFF] = output OFF (high impedance)
SEL2
Assignment of the analogue output to the process value:
[COND] = conductivity
[TEMP] = temperature
ASP2-TEMP
Analogue start point temperature; Setting range: -25...115 (°C)
hysteresis of AEP2-TEMP > 20 % of AEP2-TEMP, min. 35 (°C)
AEP2-TEMP
Analogue end point temperature; Setting range: 10...150 (°C)
hysteresis of ASP2-TEMP > 20 % of ASP2-TEMP, min. 35 (°C)
Offset-TEMP
Zero-point calibration (calibration offset) / temperature;Setting range:
+/- 5 K
ASP2-COND
Analogue start point conductivity; Setting range: 0...7500 µS/cm
hysteresis for AEP2-COND > 50 % for AEP2-COND
AEP2-COND
Analogue end point conductivity; Setting range: 100...15000 µS/cm
hysteresis for ASP2-COND > 50 % for ASP2-COND
Lo�T
Minimum value memory for the temperature
Hi�T
Maximum value memory for the temperature
Reset
[Hi�T] and [Lo�T]
Reset maximum and minimum value memory
(button to activate the system command)
Lo�C
Minimum value memory for conductivity
Hi�C
Maximum value memory for conductivity
Reset [Hi�C] and
[Lo�C]
Reset maximum and minimum value (button to activate the system
command)
FOU2
Response of OUT2 in case of a fault:
[OU] = analogue value reacts according to the process value, if
possible� Otherwise: analogue value goes to [OFF]�
[On] = analogue output switches to value > 21 mA in case of a fault\
[OFF] = analogue output switches to value < 3.6 mA in case of a fault
dAP
Damping of the measured signal; Setting range: 0...20 s
S�Tim
Simulation; enter the simulation time
Setting range: 1���60 min
S�On
Simulation; status of the simulation:
[OFF] = simulation OFF
[On] = simulation ON