37
UK
Simulation active until [Enter] is pressed again or the time set via [STim] elapses.
During the simulation [SIm] is displayed every 3 s. When the simulation has ended,
[S.On] is displayed.
The outputs react according to the simulated process values.
12 Operation
After power on, the unit is in the operating mode (= normal operating mode).
It carries out its measurement and evaluation functions and generates output
signals according to the set parameters.
►
Check whether the unit operates correctly.
12.1 Operating indicators
Numerical value + LED 1
Current level in mm.
Numerical value + LED 2
Current level in inch.
Numerical value + LED 3
Current level in % of the scaled measuring range.
LED 7 / LED 8
Switching status OUT2 / OUT1.
[----]
Level below the active zone.
[FULL] + numerical value
alternately
Level has reached or exceeded the maximum measuring
range (= overflow warning).
---- continuous
Initialisation phase after power on.
On delivery the unit is not operational. Basic settings required
(→ 10.2).
[Sim] + xxx
Simulation active. XXX = state to be simulated
[Loc]
Unit locked via operating keys; parameter setting impossible.
For unlocking press the two setting buttons for 10 s.
[uLoc]
Unit is unlocked / parameter setting is possible again.
[C.Loc]
The unit is temporarily locked. Parameter setting active via
IO-Link.
[S.Loc]
Unit is permanently locked via software. This locking can only
be removed with a parameter setting software.