10.6.4 Set output status in fault condition
► Select [FOU1] and set the value
-[On] = output 1 switches ON in case of an error�
-[OFF] = output 1 switches OFF in case of an error�
> For both values ([ON] and [OFF]) the counter stops counting in case of
a fault�
-[OU] = output 1 switches irrespective of the fault as defined with the
parameters�
► Select [FOU2] and set the value
-[On] = output 2 switches ON in case of a fault, the analogue signal
goes to the upper end stop value (22 mA)�
-[OFF] = output 2 switches OFF in case of a fault, the analogue signal
goes to the lower end stop value (3�5 mA)�
-[OU] =output 2 switches irrespective of the fault as defined with the pa-
rameters� The curve of the analogue signal corresponds to IEC60947-
5-7 (→ diagram in → 4.6).
10.6.5 Set standard pressure
► Select [rEF�P] and set the requested standard pressure�
Setting range: 13�78���15�22 psi in steps of 0�01 psi�
10.6.6 Set standard temperature
► Select [rEF�T] and set the requested standard temperature�
Setting range: 32���77 °F in steps of 1 °F�
10.6.7 Set low flow cut-off
► Select [LFC] and set the limit value�
Setting range: 0�15 %���1,05 % of the measuring range in steps of
0�15 % of the measuring range�
10.6.8 Select the medium to be monitored
► Select [MEDI] and set the requested medium:
-[Ar] = argon�
-[CO2] = carbon dioxide�
-[N2] = nitrogen�
Changing to another medium modifies the set limit values for switch-
ing signals and the analogue signal�
► Select the medium before setting the values for SPx, rPx, ASP
und AEP�