10.8 Changing the probe length
Required after a factory reset [rES] and after changes to the probe�
► Measure the probe length L to a precision of ± 2 mm (± 0�1 inch)� L =
lower edge of the of process connection to the rod end�
► Round up the measured value (step increment 0�5 cm / 0�2 inch)�
► Select [LEnG] and set the value (setting range: 10�0 ��� 160�0 cm / 4�0 ���
63�0 inch)�
11 Operation
After power on, the unit is in the Run mode (= normal operating mode)� It carries
out its measurement and evaluation functions and generates output signals ac-
cording to the set parameters�
11.1 Operating indicators
Numerical value + LED 1
Current level in cm�
Numerical value + LED 2
Current level in inch�
Numerical value + LED 3
Current level in % of the final value of the measuring range�
LED 5 ��� LED 8
Switching status of the corresponding output�
[----]
Level below the active zone�
[FULL] + numerical value
alternately
Level has reached or exceeded the maximum measuring
range (= overflow warning)�
[CAL]
Initialisation phase after power on
On delivery the unit is not operational� Entering of the probe
length is required (→ 10.2).
[Loc]
Unit electronically locked; parameter setting impossible� For
unlocking press the two setting buttons for 10 s�
[uLoc]
Unit is unlocked / parameter setting is possible again�
[S�Loc]
If [S�Loc] is displayed when an attempt is made to modify a
parameter value, either an IO-Link communication is active
(temporary locking) or the sensor is permanently locked via
software� This locking can only be removed via a parameter
setting software�