1403 Series

6

Maintenance

3.4  Permittivity Errors

The 1403 series can be used over a wide temperature 
range and barometric pressures however it is impor-

tant to understand that both aff ect the permittivity and 
therefore the capacitance.

The  change  of  humidity  from  dry  air  to  50% 
RH results in a around a 1 ppm change in permittiv-
ity and therefore a negligible change in capacitance.  

A change in atmospheric pressure from sea level to 
1600 meters, results in a change of around 100 ppm 
in permittivity and therefore a signifi cant change in 
capacitance.  

The paper The Measurement and Uncertainty of Air 
Dielectric  Capacitors  from  1  kHz  to  10  MHz  pro-
vides extremely useful information on the change of 
capacitance vs barometric pressure, relative humidity 
and temperature for the 1403 Series Air Capacitors.

https://www.osti.gov/servlets/purl/1504063

IET Labs calibrates the 1403 in it’s facility in West 
Roxbury, MA. The approximate elevation is 115 feet 
(35 meters) above sea level. Given this elevation the 
approximate air pressure is 101000 Pa. 

3.5  Design of the 1403

The 1403 is a three-terminal standard air capacitor. 
The 1403 with capacitance values < 2 pF use three 
plates to make up the capacitor element.  A high and 
low plate which  are separated by guard plate. The 
guard plate has an aperture of a specifi c diameter for 
the capacitance required.  

Higher  capacitance  values  use  a  stack  of  plates  to 
make up the capacitor element.

All plates are separated by ceramic shoulder washers 
which minimize dissipation factor.

Additional information on the design of the 1403 can 
be found at;

https://www.ietlabs.com/pdf/GR_
Experimenters/1959/GenRad_Experimenter_Aug-
Sept_1959.pdf

3.6  Environmental Conditions

3.6.1  Operating Temperature

For optimal accuracy, 1403 models should be used 
in  an  environment  of  23°C  ±1°C.   They  should  be 
allowed to stabilize at those temperatures after any 
signifi cant temperature variation.  

3.6.2  Storage Temperature

The 1403 units should be maintained within the stor-
age  temperature  range  of  0°C  to  40°C  to  retain  its 
accuracy within the specifi ed limits.

3.7  Shipping and Handling

The 1403 Series should not be exposed to any excessive 
shock or temperature extremes.  

3.8  Cleaning and Maintenance 

A soft cloth and denatured alcohol can be used for cleaning.  

A  Q-tip  and  denatured  alcohol  can  be  used  to  clean  the 
connectors.  

No other maintenance is required. 

Содержание 1403-9701

Страница 1: ...SUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC 1403 Series Capacitance Standard Operation Manual Copyright 2022 IET Labs Inc Visit www ietlabs com for m...

Страница 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Страница 3: ...t no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall n...

Страница 4: ...002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirements The a xed label indicates that you must not discard this electrical electronic product in domestic household was...

Страница 5: ...ons contained in the warnings must be followed CAUTION DO NOT Operate in an Explosive Atmosphere Do not operate the instrument in the presence of in ammable gasses or fumes Operation of any electrical...

Страница 6: ...USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working w...

Страница 7: ...ature Coe cient 5 3 4 Temperature Cycling 6 3 5 Design 6 3 6 Environmental Conditions 6 3 6 1 Operating Temperature 6 3 6 2 Storage Temperature 6 3 7 Shipping and Handling 6 Chapter 4 Maintenance 7 4...

Страница 8: ...vi...

Страница 9: ...lues ranging from 0 001 pF to 1000 pF The 1403 Standard Air Capacitors have a low di electric constant and excellent frequency response making them ideal for testing LCR meters at various frequencies...

Страница 10: ...ak 20 x10 6 1403 R 0 01 pF 0 3 20 x10 6 1403 V 0 001 pF 1 20 x10 6 Calibration An certi cate of calibration is supplied with each capacitor traceable to the SI giving the measured direct capacitance a...

Страница 11: ...m insulation and minimize the use of bare conductors b Remove power when adjusting the capacitor c Post warning signs and keep personnel safely away SN A2 2204154 1403 lbl p4 1403 K 02 05 19 Temperatu...

Страница 12: ...4 1403 Series 4 Specifications...

Страница 13: ...nnections to Capacitor All 1403 capacitors have 2 GR 874 or 2 bnc connec tors labeled HI and LO as shown in gure 3 1 Figure 3 1 1403 capacitor standard with bnc connectors Equivalent circuit showing d...

Страница 14: ...terminal standard air capacitor The 1403 with capacitance values 2 pF use three plates to make up the capacitor element A high and low plate which are separated by guard plate The guard plate has an a...

Страница 15: ...rval The recommended 1403 Series calibration interval is twelve 12 months The calibration procedure may be carried out by the user if a calibration capability is available by IET Labs or by a certi ed...

Страница 16: ...l should be maintained This connection mimics the connection on the IET GenRad 1403 Standard Capacitor 2 Determine and employ proper metrological practices Allow a con dence band for the uncer tainty...

Страница 17: ...9 1403 Series 9 Maintenance 4 4 Replaceable Parts List There are no replaceable parts on the 1403 Series Contact IET Labs at for an RMA should service be required at https www ietlabs com...

Отзывы: