5-4-5- Replacement of IDS-iSYS Wash Solution pump (Immunoassay)
•
Switch the analyser o
ff
.
•
Remove the cover from the right-hand side of the analyser.
•
Press on the two pins at each side of the pump to be changed and remove the pump body from its shaft.
•
Disconnect the pipes from the connectors and plug into the new pump.
•
Clean the shaft of the pump with absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Put the pump back in place on its shaft.
•
Replace the cover of the analyser.
•
Switch the analyser back on again and perform the start-up procedure.
•
Carry out partial priming of the washers in order to prime the tubing circuit again.
5-4-6- Replacement of IDS-iSYS System Liquid pump
•
Switch the analyser o
ff
.
•
Remove the cover from the right-hand side of the analyser.
•
Press on the two pins at each side of the pump to be changed and remove the pump body from its shaft.
•
Disconnect the pipes from the connectors and plug into the new pump.
•
Clean the shaft of the pump with absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Put the pump back in place on its shaft.
•
Replace the cover of the analyser.
•
Switch the analyser back on again and perform the start-up procedure.
•
Carry out partial priming of the arm in order to prime the circuit again.
5-4-7- Replacement of liquid waste pump
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
•
Switch the analyser o
ff
.
•
Remove the cover from the right-hand side of the analyser.
•
Press on the two pins at each side of the pump to be changed and remove the pump body from its shaft.
•
Disconnect the pipes from the connectors and plug into the new pump.
•
Clean the shaft of the pump with absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Put the pump back in place on its shaft.
•
Replace the cover of the analyser.
•
Switch the analyser back on again and perform the start-up procedure.
Maintenance
5-4-
Analyser Interventions
(continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-4
130
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...