5-3-2-2- Rinsing of the AP Substrate tubing
•
Prepare a 500 mL bottle containing de-ionised water to mimic the AP Substrate.
•
Click on the corresponding
button in front of this item in the maintenance list:
•
Replace the AP Substrate bottle with the de-ionised water bottle.
•
Click on
.
•
When the rinsing is complete, the following message displays:
•
Remove the de-ionised water bottle and replace the original AP Substrate bottle.
•
Click on
.
•
When maintenance is completed without any error, the date and time are recorded.
Maintenance
5-3- Monthly Maintenance (continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-3
125
Software version V 14