5-3-1-3- Cleaning the IDS-iSYS D-Sorb pump and level sensor shafts
!
This maintenance activity is managed only for systems using DSORB solution.
For other systems, this maintenance activity is not required.
WARNING:
OPENING/CLOSING THE LID
The lid must be opened for this operation.
Always handle the lid carefully during opening and closing.
The lid must always be fully open
and the lid support tool must always be
installed
.
Take care not to knock the lid during any analyser intervention.
When the lid is not fully open there is a risk of it falling.
During closing, maintain the lid open when removing the lid support tool, to avoid any
risk of accidental falling.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Remove the cover located at the rear.
•
Click on the corresponding
button in front of this item in the maintenance list.
•
Follow the instructions displayed on screen.
•
Release the pump body from its shaft by pinching the two clips at either side and gently pulling, ensuring the
tubes remain connected.
•
Clean the pump shaft with a cloth soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Return the pump body to its shaft.
•
Put the cover back in place.
•
Remove the lid support tool and close the lid (see page 36).
•
To clean the level sensor, use absorbent paper soaked in de-ionised water.
•
When maintenance is completed, click on
. The date and time it is carried out are recorded.
5-3-1-4- Cleaning the liquid waste level sensor
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
•
Click on the corresponding
button in front of this item in this maintenance list.
•
Remove the lid from the liquid waste bottle.
•
Clean the level sensor with bleach (commercial preparation).
•
Rinse the level sensor with de-ionised water.
•
Replace the lid.
•
When maintenance is completed, click on
. The date and time it is carried out are recorded.
Maintenance
5-3- Monthly Maintenance (continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-3
123
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...