5-2-1- General Maintenance
5-2-1-1- Cleaning the reagent compartment and Plexiglas
®
!
Do not use any spray products on the analyser
.
•
Open the reagent compartment drawer.
•
Clean the upper lid with absorbent paper soaked in de-ionised water.
•
Clean the compartment base plate with absorbent paper soaked in de-ionised water. Then use a disinfectant
wipe/solution suitable for medical devices.
•
Close the drawer.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Clean the Plexiglas
®
over the reagent compartment with absorbent paper soaked in de-ionised water. Then
use a disinfectant wipe/solution suitable for medical devices.
•
Remove the lid support tool and close the lid (see page 36).
•
When maintenance is complete, click on the corresponding
, then confirm the
maintenance by clicking on YES.
5-2-1-2- Cleaning the sample tray and Plexiglas
®
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
!
Do not use any spray products on the analyser
.
•
Open the sample drawer.
•
Clean the surface of the fixed or removable tray with a disinfectant wipe/solution suitable for medical
devices.
WARNING:
RISK OF INJURY
If a position of the fixed tray has to be cleaned, take care not to push the wipe into
the position with your finger.
Risk of cuts from the detection sensors.
Maintenance
5-2- Weekly Maintenance (continued)
IDS-iSYS User Manual - Revision M1
Maintenance 5-2
119
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...