3-14- Switching The Analyser O
ff
The analyser remains in assay mode until the run cycle is stopped.
Assays must be stopped in order to carry out maintenance procedures.
•
Click on
.
A message confirming the request displays:
•
To stop the cycle, click on
.
If assays are under way, the analyser finishes these before stopping.
Comment: The request to stop the run cycle can be cancelled by clicking on
.
•
Check that all the profiles are completed and that there are no unfinished profiles. The latter are displayed in
purple and their number is indicated on the status bar at the bottom of the screen.
•
If all the profiles are completed, open the drawer and remove all samples.
•
If reagent cartridges are not intended to be stored on-board, remove reagent racks and store in accordance
with IFU recommendations.
•
Carry out daily maintenance, then if necessary, weekly and monthly maintenance.
•
If the reagent drawer is equipped with a condensation waste, empty the drip tray if necessary.
•
Empty the solid waste tray (see
3-14-1- Emptying the solid waste
, page 101) and the liquid waste if
necessary (see
3-14-2- Emptying the liquid waste
, page 102).
•
Waste must be disposed of in accordance with current local regulations (see APPENDIX I: Waste disposal,
page A2).
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
•
Check that the work list is empty. If necessary, delete profiles that are no longer required.
•
Check the time and date of “wake-up” programmed in the SYSTEM SETTINGS menu if applicable.
•
From the Session menu, select SHUT DOWN to put the analyser into standby mode.
!
For systems using DSORB solution, a maintenance procedure for the four washers will take
place automatically during the SHUT DOWN procedure.
!
Do not turn o
ff
the computer in standby mode.
Use
3-14-
Switching The Analyser O
ff
IDS-iSYS User Manual - Revision M1
Use 3-14
100
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...