3-10-3- Sample results
•
Click on the position occupied by the sample.
•
The results obtained are displayed with any associated messages.
•
In the case of multiple replicates, the last result is displayed on a yellow contour.
•
Double-click on the result to view the results for all replicates.
•
To confirm a result, an analysis can be repeated. To do this, check the box in front of the analyte to be
repeated.
•
The results can be printed out.
If the value cannot be calculated or is outside the field of measurement, the result is replaced by
* * * * *
.
* * * * *
is always accompanied by a message (see
3-11- Messages Associated with Results
, page 89).
If the value is outside the field of measurement, the following information is displayed:
-
value below the field of measurement: the message OMR- is accompanied by «< Low limit of the
sample measurement range». When the field of measurement starts at 0, the result is replaced by * * *
* *.
-
value above the field of measurement: the message OMR+ is accompanied by «> High limit of the
sample measurement range».
Limits are expressed in working units.
Comment:
When the message ORA accompanies the messages OMR- or OMR+, the result is not calculated and is
replaced by
* * * * *
.
If the analyser is connected to a centralised computer system, the information sent for results
outside the field of measurement depends on the selected connection protocol:
•
ASTM Compatible: results are always sent in the format * * * * * ..
•
ASTM V2 Compatible and ASTM V3 Compatible: results are sent in the following format:
-
OMR-
: < Low limit of the sample measurement range, even when the measuring range
starts at 0.
-
OMR+
: > High limit of the sample measurement range.
Limits are expressed in working units.
!
Programming reflex tests
When reflex tests can be launched for a sample, the button
is displayed.
•
Click on
to program the reflex tests.
The new analyse requests are added to the profile.
Comment: an analysis requested by a reflex test can be removed by unchecking the analysis.
•
Then click on
.
Use
3-10-
Results (continued)
IDS-iSYS User Manual - Revision M1
Use 3-10
86
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...