MSD-62 Series User’s Guide
22
7 Precautions
Before connecting to external devices, follow the precautions below.
7.1 Installation
When installing the MSD, please observe the following precautions.
● Do not stack or place one MSD directly on top of another MSD.
● Do not block vent holes. To provide adequate ventilation, maintain sufficient clearances around the
MSD (30 mm/1.18 inches or more).
● When the MSD needs to be mounted in an EIA rack, or an enclosed space, ensure that sufficient
ventilation or cooling is provided and that the ambient temperature will not exceed 40°C / 104°F. If
inadequately vented, the product’s service life, operation and reliability may be affected.
■
Attaching rack mounting brackets
1.
Remove four M5 screws from one side of the unit. Retain these screws for step 2.
2.
Attach one bracket to the side of the unit using the screws removed in step 1.
3.
Repeat steps 1 and 2 on the other side of the unit.
[Fig. 7.1] Attaching rack mounting brackets
Note:
The standard screw tightening torque is 2.94 N
・
m (about 30 kgf
・
cm).