14 Troubleshooting (problem/possible cause/solution) 

Send any questions about the Idiag P100 and its accessories or equipment to the address on the last page 

of this user manual. To allow immediate processing and a faster response, please provide a complete 

return address and the serial number of the Idiag P100 in question. 

 

Problem 

Possible cause 

Solution 

The device does not turn on 

Low battery level 

Charge the battery and try again 

The device does not turn on 

Battery not installed 

correctly 

See Chapter 9.1. Ensure that the battery is 

installed with the contacts the right way 

around. 

The device does not turn on 

The detachable 

module is not 

correctly fitted 

Place the detachable module with its three 

struts at a 45-degree angle to the base 

module and then bring the two halves of the 

grip together. Once it has been correctly 

assembled, the unit will beep.  

The device does not turn on or off or the 

training intensity cannot be changed. 

Defective 

button/device 

crashed 

Remove the battery and reinstall it. Re-start 

the device. If the problem persists, return 

the device to Idiag after contacting us by 

telephone (see address on last page). 

Menu navigation does not work as 

desired. 

Software error 

Turn the device off, remove the battery, 

reinstall it and turn the device on again. If 

the problem cannot be resolved, check the 

Idiag P100sport App to see if there are any 

software updates available.  

The air guide is not recognised 

 

Air  guide  not fitted 

correctly 

Remove the air guide from the base module 

and refit it. The air guide may need to be 

pressed on with some force to ensure the 

contacts connect properly. Ensure that the 

air guide is completely dry after washing it. 

The base module and detachable 

module cannot be properly connected 

The detachable 

module is not 

properly engaged 

Remove the detachable module and refit it. 

Ensure that the top Chapter  is fitted first, 

before connecting the handles.  

The battery does not charge  

Battery not installed 

correctly 

Check that the battery is correctly installed 

such that the contacts are correctly 

positioned in the charger.  

Also check that the mains plug is properly 

inserted into the charger.  

Breathing bag is leaking 

The breathing bag is 

not correctly rolled 

up 

Ensure that the breathing bag is correctly 

rolled up and sealed before connecting it to 

the device.  

The mouthpiece loses its shape 

Normal wear and 

tear 

The mouthpiece is a wear part. If the 

mouthpiece loses its shape, we 

recommend that you purchase a new 

mouthpiece to ensure the best possible 

training.  

Immediately after switching on, a 

warning signal sounds and the device 

switches off 

Air guide not properly 

recognised 

After cleaning, allow the air guide  to dry 

completely (inside and out), reinsert it and 

restart the device. Alternatively, check that 

the detachable module has been correctly 

inserted into the base module. 

 

 

15 Disposal 

 

This product, including its associated battery, must not be disposed of together with normal 

household waste. Before disposing of this product please consult your local authority or 

your dealer and enquire about the correct method of disposal.  

 

 

 

 

 

 

Содержание P100

Страница 1: ...EN IDIAG P100 User manual Version 1 0...

Страница 2: ...duction 9 5 Indications 9 6 Contraindications 9 7 Adverse effects 9 8 Risk factors 9 9 Before operating the Idiag P100 10 9 1 Charging installing and removing the battery 10 9 2 Assembling the P100 11...

Страница 3: ...e conditions 27 13 Technical specifications 28 14 Troubleshooting problem possible cause solution 29 15 Disposal 29 16 Warranty 30 16 1 Warranty conditions 30 16 2 Warranty coverage 30 16 3 Legal prov...

Страница 4: ...in advance by reading the user manual 1 3 Instructions for use This user manual is also available in PDF format Hyperlinks and bookmarks allow links to be opened quickly Moreover the search function f...

Страница 5: ...dition The use of the device is limited to the predefined purpose and application as described in this user manual See also Chapter 12 4 Storage conditions The underlying principle of the Idiag P100 i...

Страница 6: ...ea or vestibular disorders In such cases training must be stopped immediately The user carries full responsibility for all types of training application Idiag assumes no responsibility for the trainin...

Страница 7: ...ghtly from those shown 3 2 Additional parts The Idiag P100sport App can be downloaded for iOS and Android from the respective App Stores Links to the current version of the App can be found on our web...

Страница 8: ...tend to train their respiratory muscles However none of the contraindications listed below may be present 6 Contraindications Age 6 years Doctor s recommendation to avoid physical stress Pregnancy Acu...

Страница 9: ...e may only be used with the supplied battery and associated charger Remove the battery from the device before charging it Remove the battery from the device if it is not to be used for a longer period...

Страница 10: ...the inserted air guide 3 and the detachable module 2 with the inserted battery 1 To do this place the detachable module 2 at a 45 degree angle flush to the base module 4 and bring the two halves of t...

Страница 11: ...Results Training settings 9 4 Turning the device on and free training When the device is turned on for the first time the available training modes appear for selection If all of the training modes do...

Страница 12: ...ore training for the first time or as necessary you can take the assessment test to automatically determine your training settings The parameters can also be manually adjusted at any time 9 5 Starting...

Страница 13: ...d strength Training can be started by tapping START or you can return to the training mode selection screen by tapping LATER 5b Tap START to begin training with the previous test results If no test re...

Страница 14: ...n to the original values Tap START to begin training step 9 Tap the arrow to return to step 7 8b Change the breathing resistance by tapping Tap RESET to return to the original values Tap START to begi...

Страница 15: ...7 10b Training completed successfully Tap RESTART to start training again with the same settings step 6 Tap FINISH to go to the training mode selection screen Tap TRAINING RESULTS to view the results...

Страница 16: ...tory muscle endurance test has been successfully concluded the training settings will be configured in accordance with the test results Training can be started by tapping START or you can return to th...

Страница 17: ...the breathing frequency tap FREQUENCY step 7b To set the depth of respiration tap DEPTH step 7c Tap the arrow at the bottom edge of the screen to return to step 4 7a Change the duration of training by...

Страница 18: ...inue training without any changes Tap STOP to stop training and go to the currently selected training mode step 5 Select TRAINING SETTINGS to change the current training settings step 6 9b Training co...

Страница 19: ...uency shows you how many breaths you took per minute during training The duration shows you for how long you trained in minutes and seconds The power of breathing shows you an average value of the ene...

Страница 20: ...Breathing depth OK Breathe shallower Breathe deeper Breathing frequency End of expiration Middle of inspiration End of inspiration Middle of expiration End of expiration Note For additional respirator...

Страница 21: ...ting tone if the depth of respiration is too shallow or a calming tone if the depth of respiration is too deep 10 3 Visual feedback during strength training Threshold value achieved successful breath...

Страница 22: ...hpiece for cracks and leaks Wear parts particularly the breathing bag and mouthpiece must be regularly replaced for hygiene and safety reasons according to the information in Chapter 13 Technical spec...

Страница 23: ...ean The air guiding parts are not intended for use by different users Every 2 3 weeks Clean Clean and disinfect Replace See Technical specifications on page 28 11 4 Cleaning Frequency See table in the...

Страница 24: ...e surfaces with a lint free cloth dampened with water and mild soap Rinse air guiding parts by hand Allow all cleaned parts to dry thoroughly before using them again for training and before storing th...

Страница 25: ...s Never immerse the base module or detachable module in disinfectant The use of a spray is not recommended because the disinfectant could penetrate the mechanical parts of the Idiag P100 If a disinfec...

Страница 26: ...Remove the battery from the Idiag P100 if you are not going to use the device for a period of several days or longer 12 2 Operating conditions Temperature 10 C to 40 C 50 F to 104 F Relative humidity...

Страница 27: ...ontact with flammable mixtures Mode of operation Continuous Information regarding faults and fault prevention This device has been tested and complies with the EMC Directives and the Medical Device Di...

Страница 28: ...itted correctly Remove the air guide from the base module and refit it The air guide may need to be pressed on with some force to ensure the contacts connect properly Ensure that the air guide is comp...

Страница 29: ...ervice organisation authorised by the manufacturer may perform any necessary repairs o Accidents force majeure or other causes beyond Idiag s control such as water damage fire unauthorised modificatio...

Страница 30: ...RLAND Idiag AG M listrasse 18 CH 8320 Fehraltorf T 41 0 44 908 58 58 F 41 0 44 908 58 59 E info Idiag ch idiag ch GERMANY Idiag GmbH Romanstrasse 72 D 80639 Munich T 49 89 2000 297 20 E info Idiag de...

Отзывы: