Operator Manual ilex with IPS500A
L280-0688 REV C
13.
ME Equipment and Systems
Table 1 per 5.2.2.1.c of IEC 60601-1-2: 2007 for CISPR 11 ME Equipment and ME Systems.
Guidance and manufacturer’s declaration –electromagnetic emissions
The ilex system is intended for use in the electromagnetic environment specified below. The customer or the user of the ilex system should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance.
RF emissions CISPR 11
Group 1
The ilex system uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class A
The ilex system is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-
3-2
Class A
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
Table 2 per 5.2.2.1.f of IEC 60601-1-2: 2007.
Guidance and manufacturer’s declaration –electromagnetic immunity
The ilex system is intended for use in the electromagnetic environment specified below. The customer or the user of the
ilex system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power supply
lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the ilex system
requires continued operation
during power mains
interruptions, it is
recommended that the ilex
system be powered from an
uninterruptible power supply
or a battery.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
3 A / m
3 A / m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.