Operator Manual ilex with IPS500A
L280-0688 REV C
Linak
website.
Battery (12V/1.2Ah)
Part Number K000-0040 (Replace in pairs)
Note:
The battery will not charge if the system is not plugged in or the green toggle
switch/thermal breaker is off.
8.
General Cleaning
Tools: Soft cloth and approved cleaner:
o
Sodium hypochlorite (generic household bleach) in a solution of 5.25% sodium hypochlorite
diluted between 1:10 and 1:100 with water.
o
Alcohol (generic).
o
Envirocide. ™
Notes:
•
For the cleaning of monitor or shelf equipment, please consult corresponding literature
for these devices.
•
NEVER use alcohol to clean the monitor.
Procedure:
1.1.
Unplug the ilex
system prior to the cleaning process.
1.2.
Wipe clean the surfaces of the system with a diluted mixture of mild detergent (listed above)
and water, using a soft cloth.
Use a soft cloth to dry any stainless-steel items.
1.3.
To clean speaker grilles, simply pull off and wash.