Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
15
8.5.
FOOT ACTUATED MOTION CONTROLS
100-4T G3 FOOT SWITCH CONTROL
Foot Switch activated control motions for the
Aspect
100-4T G3 Table Top are: elevate/lower
and Trendelenburg down/up. (Safety hood helps
avoid accidental motorized table movement
from Foot Switch).
•
When the Trendelenburg Foot Pedal is
used, the tabletop motion will stop at level
position for 1 second before motion begins
again. When the Table’s Trendelenburg
position is already level and the Trendelenburg
Foot Pedal is pressed, it will take 1 second
before the motion of the Table Top begins.
ISR G3 FOOT SWITCH CONTROL
Foot Switch activated control motions for the
Aspect
ISR G3 table are: ISO-Roll,
Trendelenburg down/up and elevate/lower.
(Safety hood helps avoid accidental motorized
Table movement from Foot Switch Control).
•
When the Trendelenburg Foot Switch Pedal
is used, the Table Top motion will stop at
level position for 1 second before motion
begins again. When the Table’s
Trendelenburg position is already level and
the Trendelenburg Foot Switch Pedal is
pressed, it will take 1 second before the
motion of the tabletop begins.
•
When the ISO-Roll Foot Switch Pedal is
used, the Table Top motion will stop at
level position. When this happens, release
and repress the ISO-Roll Foot Switch Pedal
again to continue the motion.