Operator Manual
Aspect 100-4
L100-3316 Rev. A
24
13.
TROUBLESHOOTING
Note: The motion of the Tabletop is fully controlled by user interface with switches,
buttons, and a hand actuated brake release. In the event of a loss of these motions it is
expected that the Tabletop will remain stationary without any unwanted movement.
Problem/Symptom
Possible Cause
Remedy
1. Table controls
are not functioning.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Control connection.
5. Electromagnetic
Interference.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Service Controller connections
according to Service Manual.
5. Refer to Section 5 of this manual.
2. Lateral and
Longitudinal Brakes
stuttering or
sticking.
1. Table has been dormant
for extended length of
time.
2. Contamination on brake
surfaces, or improper
cleaning or maintenance
performed.
1. Release brakes using Tabletop
Motion Control Panhandle and move
tabletop in all directions. Working
tabletop motions may free dragging.
2. Clean brake surfaces according to
Service Manual.
3. Lateral and
Longitudinal Brakes
not releasing.
1. No Power.
2. Emergency Stop
Pushbutton engaged.
3. Battery is depleted.
1. Check electrical outlet.
2. Reset Emergency Stop Pushbutton.
3. Connect to AC Power Outlet.
4. Longitudinal
Brakes not
releasing.
1. Emergency Stop
Pushbutton engaged.
1. Reset Emergency Stop Pushbutton.
5. Table does not
elevate.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Panhandle Failure.
5. Control connection.
6. Actuator Failure.
1. Check electrical outlet.
2. Connect to AC Power
3. Reset Emergency Stop Pushbutton.
4. Test motion with Handset Control. If
elevate works correctly, replace
Panhandle Unit
5. Service Control connections
according to Service Manual.
6. Replace Elevate Actuator according
to Service Manual.