background image

10

Filta-Max 

xpress 

Operator’s Guide

Setting up the Pressure Elution Station

Location

The Filta-Max 

xpress

 Pressure Elution Station and buffer reservoir bottle should be placed on a 

sturdy bench or laboratory surface capable of supporting both their weight and the force required to 
install and remove the Pressure Elution Station housing and fittings.

Power connection

•  Connect the low-voltage cable to the inlet power socket on the back panel of the Pressure 

Elution Station.

•  Connect the provided power supply to a suitable electrical outlet using a power cord with 

the appropriate rating (18 AWG/.75 mm

2

). The external power supply will function normally 

between 100 and 240 V AC at 47–63 Hz.

Connecting the air supply

Compressed air is required for the Pressure Elution Station to function properly because it supplies 
the air pressure needed to elute the Filter Modules. IDEXX supplies a hose kit that lets you connect 
either your facility’s air supply or a compressor to the Pressure Elution Station. The hose kit consists 
of a 12-foot self-retracting hose and connector.
Compressed air must be oil and moisture free. IDEXX can supply an external air filter that attaches 
to the rear of the Pressure Elution Station to remove oil and moisture from the air.

Using your facility’s air supply

If you use your own facility’s air supply, check to ensure that the outlet pressure is between 5.0 bar 
and 8.0 bar (72.5 and 116 psig). The air supply must have a sufficient capacity to perform elution 
cycles whilst remaining above 5.0 bar (72.5 psig). Your facility’s air supply may require filters to 
remove oil and moisture. 
Use the quick-connect fittings supplied with the Pressure Elution Station or other appropriate fittings 
to connect the air supply. The connection is made at the back of the Pressure Elution Station at the 
air line inlet connector.
Before connecting and applying air pressure to the Pressure Elution Station, ensure that the 
lockable air inlet valve is in the OFF (EXH) position by turning it clockwise. When the connection 
has been made and there is pressure to the Pressure Elution Station, turn the lockable air inlet valve 
counterclockwise to the ON (SUP) position.

Connecting an air compressor to the hose kit

Note:

 The compressor must be oil free.

1.  Prepare the air compressor for use according to its operating instructions. 

Note: 

The components and their respective locations on the air compressor may vary with 

the model. Polytetrafluroethylene (PTFE) tape, plumbing tape, or other appropriate sealant is 
recommended at all threaded connections.

2.  Connect the swivel end of the 12-foot self-retracting hose (found in the Filta-Max 

xpress

 

Hose Kit) to the air compressor’s filter, which is attached to the regulator.

3.  Ensure that the lockable air inlet valve on the Pressure Elution Station is in the OFF (EXH) 

position by turning it clockwise.

4.  Connect the other end of the hose to the air line inlet connector on the Pressure Elution 

Station.

Содержание Filta-Max xpress

Страница 1: ...Filta Max xpress Pressure Elution Station Operator s Guide...

Страница 2: ...e patents or pending patent applications trademarks copyrights or other intellectual or industrial property rights covering this document or subject matter in this document The furnishing of this docu...

Страница 3: ...connection 10 Connecting the air supply 10 Using your facility s air supply 10 Connecting an air compressor to the hose kit 10 Checking the pressure regulator setting 11 Assembling sampling and filter...

Страница 4: ...n 20 On a daily basis 20 On a monthly basis 20 Servicing the Pressure Elution Station 21 Storing the Pressure Elution Station 22 Limited warranty 24 Appendix A Consumables 25 Appendix B Air supply req...

Страница 5: ...een the Filter Housings and any sampling equipment are capable of withstanding the operating pressure of the elution station The Filta Max xpress Pressure Elution Station weighs approximately 86 lb 38...

Страница 6: ...damaged or leaking unit Always turn off the power and disconnect the power cord prior to performing any service or routine maintenance Never operate the Filta Max xpress Pressure Elution Station with...

Страница 7: ...n vessel holder Figure 1c Sensors fittings and vessels Figure 1 Pressure Elution Station without a Filter Housing and diverter assembly Socket head cap screw Pressure chamber lid Buffer inlet with flo...

Страница 8: ...r pack Filta Max xpress Operator s Guide Pressure chamber O ring Tool kit 2 hex wrenches quick connect fitting screwdriver Figure 2 Pressure Elution Station with a Filter Housing and diverter assembly...

Страница 9: ...t supplied by IDEXX Sampling pump that is suitable for use with the sampling fittings or a suitable sampling rig for use with a pressurized water supply Platinum cured tubing 13 15 mm internal diamete...

Страница 10: ...e is between 5 0 bar and 8 0 bar 72 5 and 116 psig The air supply must have a sufficient capacity to perform elution cycles whilst remaining above 5 0 bar 72 5 psig Your facility s air supply may requ...

Страница 11: ...according to the manufacturer s instructions 10 When you have attached the quick connect fittings to the Pressure Elution Station and there is pressure turn the lockable air inlet valve counterclockw...

Страница 12: ...ed upstream of the Filter Housing Note A hose clip is required to ensure that the seal between the hose and the sampling inlet is watertight 3 Connect the Swagelok quick connect QC stem to a Swagelok...

Страница 13: ...n Station In this case the Filter Module must be transferred to a Filter Housing suitable for use with the Pressure Elution Station For more information refer to Eluting the Sample Operating the Press...

Страница 14: ...are generated during sample elution This initial step also serves to purge reagent grade water from the lines and replace it with buffer solution 1 Place a blank Filter Module into a Filta Max xpress...

Страница 15: ...press and hold the F2 button for 3 or more seconds Listen for any air leaks and monitor the low pressure gauge for pressure changes F3 resets the cycle counter Press and hold the F3 button for 3 secon...

Страница 16: ...lter Housing QF stem facing upward attach the outlet diverter fitting to the QF stem by pulling back the collar and pushing the fittings together 4 Turn the Filter Housing over and place the outlet di...

Страница 17: ...upport cushions for 500 mL centrifuge tubes In good working order with bucket supports lightly greased Centrifuge setup Locate centrifuge to minimise vibration Set RCF to 2000 g Set acceleration to ma...

Страница 18: ...50 mL centrifuge tubes In good working order with bucket supports lightly greased Centrifuge setup Locate centrifuge to minimize vibration Set RCF to 1500 g Set acceleration to maximum and decelerati...

Страница 19: ...ater Note If sample elution is performed infrequently during a day for example early morning and late afternoon it is advisable to perform a buffer purge between elution events This will increase the...

Страница 20: ...finished water and then with reagent grade water Relubricate and or replace all O rings as required On a daily basis Clean the surfaces of the Filta Max xpress Pressure Elution Station with laborator...

Страница 21: ...Replace the four socket head screws and tighten them using the large hex wrench 8 Replace the top lid of the Filta Max xpress Pressure Elution Station and tighten the screws using the small hex wrenc...

Страница 22: ...and silicon tubing before storage Note Do not detach the closure cap from the air and blue liquid feed lines Instruments with a touch screen control panel A Filter Housing without a Filter Module an...

Страница 23: ...e closure cap 3 Attach a Filter Housing and diverter to the Pressure Elution Station and place a centrifuge tube in the holder 4 Perform three elution cycles and check for any errors or issues with fu...

Страница 24: ...nated facility for examination and testing IDEXX shall repair or replace within 30 days of receipt by IDEXX any such product found to be so nonconforming and promptly return such products to Buyer tra...

Страница 25: ...length 5 5 cm 2 2 in diameter Storage 16 C 24 C 60 F 74 F Filta Max Filter Module black caps Materials of construction open cell reticulated polyurethane rings compressed to form a filter Ring size 5...

Страница 26: ...hose kit consists of a 12 foot self retracting hose and connector Note The air supply should be set to a minimum pressure of 5 0 bar 72 5 psig and a maximum of 8 0 bar 116 psig For details regarding...

Страница 27: ...oir bottle is sealed correctly and is airtight There is sufficient buffer solution in the buffer reservoir All tubes are attached correctly and are airtight High Flow error message is displayed and th...

Страница 28: ...ort idexx com China 86 21 61279528 Japan 81 422 71 5921 Australia 1300 44 33 99 For countries not listed above visit our website at idexx com water References 1 Drinking Water Inspectorate The enumera...

Отзывы: