background image

 

 

 
 

 

IDEAL INDUSTRIES, INC. 
TECHNICAL MANUAL 
MODELS: 61-764 

61-766 
61-768

 

 
 
The Service Information provides the following information: 

   

  Precautions and safety information 

•   Specifications     
•   Performance test procedure 
•      Calibration and calibration adjustment procedure 
•      Basic maintenance (replacing the battery) 
 

 

 

Form number:  TM61764-6-8 
Revision: 5. Date:    Nov 2007

 

 
Form number TM61764-6-8 

Rev 5 November 2007 

Содержание 61-764

Страница 1: ...ovides the following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery F...

Страница 2: ...fications and Compliance 3 Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 4 Resistance Specifications 4 Frequency Specification 4 Capacitance Specifications 4...

Страница 3: ...g this product In this manual a Warning identifies conditions and actions that pose hazard s to the user a Caution identifies conditions and actions that may damage the meter or the test instruments T...

Страница 4: ...duct only as specified CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practice...

Страница 5: ...ty 0 C to 50 C 32 F to 122 F 0 70 RH Storage Environment 20 C to 60 C 4 F to 140 F at 80 relative humidity Power source 9V Battery NEDA 1604 Battery Life 200 hours typical alkaline 61 764 200 hours ty...

Страница 6: ...6 600V 66 00V 660 0V 1000V 0 5 2 digits 66 00A 660 0A 50Hz 60Hz 1 7 10 digits AC Current True RMS 66 00A 660 0A 60Hz 400Hz 3 0 10 digits Capacitance 6 600 66 00 660 0 F 3 0 10 digits Frequency 20Hz 4...

Страница 7: ...es under 7 5V DC replace the battery see Battery Replacement page 11 before beginning the performance test 3 Input the values listed in Table 1 through 7 Table 1 AC Voltage Test Function Range Input L...

Страница 8: ...768 Table 5 Resistance Test Function Range Input Low Limit High Limit Model Number 660 100 98 6 101 4 61 764 61 766 61 768 6 6K 1K 986 1 014 61 764 61 766 61 768 66K 10K 9 86 10 14 61 764 61 766 61 7...

Страница 9: ...uracy that exceeds by an expectable ratio the accuracy of this instrument V DC Calibration 61 764 Refer to Figure 1A 61 766 Refer to Figure 2A 61 768 Refer to Figure 3A 1 Set the function range to 6 6...

Страница 10: ...ero Calibration 61 768 Refer to Figure 3B 1 Set the function range to A DC 2 Short the V and COM inputs on the meter 3 Adjust VR5 VR 5K until display reads 000 4 De energize source and remove test lea...

Страница 11: ...Page 9 61 764 Figure 1A 61 764 Figure 1B 61 766 Figure 2A 61 766 Figure 2B Form number TM61764 6 8 Rev 5 November 2007...

Страница 12: ...Page 10 61 768 Figure 3A 61 768 Figure 3B Form number TM61764 6 8 Rev 5 November 2007...

Страница 13: ...lips head screwdriver to remove the screws on battery cover 3 Remove battery from the battery compartment 4 Install new 9V battery NEDA 1604 An alkaline type is recommended 5 Install new battery into...

Отзывы: