CHAPTER 3
STRUCTURED CABLE FIELD
CALIBRATION AND TESTING
FIELD CALIBRATION – CAT3/5E/6/6A AND CLASS C/D/E/Ea/F/Fa
(TERA AND EC 7)
This Field Calibration is a 4-step process. Steps 1 and 2 are performed with the
patchcords connected to the Handsets. Steps 3 and 4 are performed with open-ended
patchcords (Only one end connected to the Handsets).
To calibrate the tester, perform the following:
Step 1
¾
Connect the adapters to the Display Handset (DH) and Remote Handset (RH).
¾
Power both units on.
¾
Connect the patchcord that you plan to use as the RH patchcord to the adapters
of the DH and RH units.
¾
From the DH Ready screen, select Field Calibration. The Calibration screen
appears.
¾
From the DH Field Calibration screen, select START to begin the calibration
process on the first (RH) patchcord. This first process takes about 10-20 seconds
to complete.
¾
At the completion of the first calibration process, tag the RH end of the first
patchcord. Disconnect the first patchcord from the DH and RH unit adapters.
This tag will remind you which end to reinsert into the RH for Step 4.
Step 2
¾
Insert the second (DH) matching patchcord into both the DH and RH adapters.
¾
From the DH Field Calibration screen, select START to begin the second
calibration step.
¾
At the completion of the second patchcord step, disconnect one end of the
second patchcord from the RH unit adapter (leaving the second end still attached
to the DH unit).
3-1