9.2.2
Input (INPUt)
Use these commands to change parameters related to the inputs. You can replace # by the numbers
1 to 4 to refer to the input that is so labelled on the device’s front panel.
Additional information about the SELECt function:
•
UNSHaped: The received signal from the physical input port is transmitted by the input block
for further processing without any change
•
SHAPed. The selected edge (rising or falling) of the received signal from the physical input port
is considered as an event. This event is transmitted further by the input block by a 4ns pulse.
•
OUTPut: The input block receives the data from the output block of the same label, e.g.,
INPUt3 receives the signal from the OUTPut3.
Input block
INPU#
(# = 1...4)
Command tree
Input
Output
Description
INPUt#
Select input number
:ENABle
{ON|OFF}
Enable/disable input
:ENABle?
{ON|OFF}
:COUNter?
number
Returns the number of counts over the
integration time (CYCLE mode) or since
last :RESEt (ACCUM mode)
:INTEgrationtime
time in ms
Set counter integration time (only in
CYCLE mode)
:INTEgrationtime?
time in ms
Returns integration time
:MODE
{CYCLE|ACCUM}
Select the counter mode
:MODE?
{CYCLE|ACCUM}
Returns the counter mode
:RESEt
Reset counter to 0
:COUPling
{AC|DC}
Set AC or DC coupling
:COUPling?
{AC|DC}
:EDGE
{RISIng|FALLing}
Set discriminator edge
:EDGE?
{RISIng|FALLing}
:THREshold
value in V
Discriminator threshold
:THREshold?
value in V
:HIREs:ERROr?
error
code
Return last high resolution error code
:HIREs:ERROr:CLEAr
Acknowledge error
:SELEct
{UNSHaped
|SHAPed
|OUTPut
|LOOP}
Select what feed the input block. For
details,
refer
to the information below
this table.
:SELEct?
{UNSHaped
|SHAPed
|OUTPut
|LOOP}
Returns what feed the input port
:STATe?
State of the block