SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 12
Technical Service Manual
5.2.10
RINSING
THE
INFUSER
To rinse the infuser after cleaning use the rinse time interval indicated in
proceed as follows:
1. Dampen a lint-free cloth with tap water.
2. Wipe each surface with the moist cloth for the entire rinse time interval.
3. For hard to reach areas, dip the swab into the water and press the tip against the
container to remove excess water and rinse the areas.
Note:
Do not allow fluid into any ports or vents.
5.2.11
DRYING
THE
INFUSER
To dry the infuser after cleaning refer to the drying time indicated in
and proceed
as follows:
1. Allow the infuser and accessories to air dry.
2. Inspect all surfaces and recessed areas for moisture.
3. When no moisture is detected, close the vial compartment door.
Note:
Allow to air dry completely before beginning disinfection.
Note:
After the vial compartment door receptacle is dry, close the vial
compartment door to avoid damage when moving the infuser.
5.2.12
CLEANING
SPILLS
DURING
THE
COURSE
OF
PATIENT
CARE
The following procedure describes how to clean non-hazardous spills or soil from the
infuser during the course of patient care.
- Non-hazardous fluid spills should be wiped up as soon as possible, and not allowed
to dry on the infuser.
- Hazardous spills (such as blood or chemotherapy drugs) should be processed per
facility policy.
To clean non-hazardous spills or soil at the patient site:
1. Inspect the infuser enclosure, display, and keypad for visible cracks or damage that
may allow fluid to reach internal components.
2. With gloves on, remove a wipe from the dispenser and unfold it to expose the
maximum surface area before wiping, or spray an approved cleaning solution on a
clean, lint-free cloth. (See
.)
CAUTION:
DO NOT USE THE INFUSER IF THE ENCLOSURE, KEYPAD, OR
DISPLAY IS DAMAGED OR CRACKED. IF DAMAGE OR CRACKS ARE FOUND,
REPLACE THE INFUSER.
CAUTION:
DO NOT SPRAY CLEANING SOLUTIONS TOWARD ANY
OPENING IN THE INFUSER.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...