1.10 INFUSION SYSTEM INSTALLATION
Technical Service Manual
1 - 9
LifeCare PCA
Figure 1-2. Setting the Time and Date
1.10.5
SETTING/CHANGING
THE
ALARM
VOLUME
The infuser is shipped with its alarm volume set to High, which is its maximum alarm
volume level.
To set the default alarm volume, proceed as follows:
1. Confirm the AC power indicator on the front panel is lit.
2. Unlock and open the security door.
3. Enter
Biomed Mode
by pressing and holding
[ENTER]
and
[ON/OFF]
simultaneously.
4. Verify the
BIOMED MODE
screen displays.
5. Verify the infusion system is not connected to a patient and press
[CONTINUE]
.
6. After the power up RAM test is completed, the infuser displays the
Biomed Mode
Main Menu
screen. Press the
[MAINTENANCE]
softkey.
7. Press the
[SET VOLUME]
softkey.
WARNING: CHECK THAT THE ALARM VOLUME LEVEL IS APPROPRIATE FOR THE
CURRENT PATIENT AND BACKGROUND NOISE LEVEL BEFORE USE ON THE
PATIENT.
CAUTION:
Setting the alarm volume to Low will result in an audible alarm
volume below the 45 dBA minumum specified by IEC60601-2-24.
SELECT
SETTING
TO CHANGE
CONTRAST
TIME/DATE
CONTINUE
SELECT TIME
WITH NUMBER
BUTTONS
TOGGLE AM/PM
10:43 PM
AM/PM
NEXT
WARNING!
CHANGING THE
DATE OR TIME
WILL CLEAR ALL
TOTALS!
CONTINUE
PREVIOUS
SET DATE WITH
NUMBER
BUTTONS
NEXT
SELECT
SETTING
TO CHANGE
04/14/12
9:21 PM
04/14/12
TIME/DATE
CONTRAST
CONTINUE
CONFIRM
CURRENT
TIME AND DATE
SETTINGS
CONFIRM
CHANGE
CANCEL
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...