manualshive.com logo in svg
background image

IDM

-60

R    R

s485

1. NOTICE

1. Please check the packaging and appearance after  you received  the 

product.
2.Please  check  lhe  mode  and  specification  of  the  product.  Keep 
correct electrical connection and mounting method. 

3.Please installation WITHOUT power.
4.During  using,  please  notice  the technical and  operation condilions:

the temperature of medium, burst pressure, power supply.

5. This product is a precision device, DO NOT use hard objects to tauch

the diaphragm or disassemble. 

6. During installation, do not install or disassemble strongly, especially 

the installation of thread. 
7. Please install or disassemble with a suitable wrench, Otherwise, the 
damage is not in the scope of warranty. 

8. After  starting  the  gauge,  it  is  a  normal  phenomenon to take  a  few 

minutes to keep it stable and work normally. 
9.lf  the product is out of order, please contact our after-sale technical 

staff.
1 0.During installation, the gauge will be affected by installation stress.

After installation, please  zero clearing.

(!!)The  product  damage  caused  by  non-professional  operation 

according to the operation regulation is not the scope by the warranty. 

2. WARNING

1.When  operation temperature is above  60 degree,  please  use forced 
fans  or  cooling  machines,  but  do  not  let  the  cooling  air  blow  to the 

switch directly. 
2.The  installalion,  commissioning  and  maintenance  of  this  product 
shall be conducted by qualified engineers or technicians member. 

3.lf  the  malfunction  or  exception  of  this  product is likely to result in a 
major  system  accident,  please  sei up an appropriate  protective  circuit 

outside to prevent accidents. 
4.The  company  does not  undertake  any  direct  or indirect  lass  other 

than the product itself.
5.The company reserves the right to change the manual of the product

without notice. 

3. DESCRIPTION

This  digital  communication  pressure  gauge  is  a  high-precision 

intelligent digital remote  pressure  gauge based  on  MODBUS  protocol 

RS485 output. 

The  digital  communication  pressure  gauge  adopts  a  4  digit  LCD 

screen, which can display real-time pressure on the  spot. At  lhe same 

time, the pressure signal is transmitted remotely,  and the transmission 

distance is better than 1000 meters. 

Built-in  pressure  sensor,  powerful  function,  users  can  change 

pressure uni!, calibrate zero deviation, correct filter constant, etc. 

This digital communication pressure gauge adopts SS 304 housing 

and connection,  waterproof cable outlet,  waterproof rating of  IP65, and 

a wide range of media suitability. 

* With RS485 signal output 

* Adopt standard MODBUS RTU communication protocol 

* Adopt all 304 stainless steel shell,  strong and durable 

* High precision, lang transmission distance, stable performance 

* Compact appearance, diameter 60mm

* Compact appearance, diameter 60mm

*High degree of intelligence,  users can sei the address, baud rate, 

filter constant, re-check, etc. 

4. TECHNICAL SPECIFICATIONS

Pressure range 

Vacuum range:  (0--100)kPa 
Compound range: -0.1-0.1 ... 0.25 0.6 1.0 1.6 2.5MPa 
Micro pressure range: -100 ... -25 ... -5-2 ... 5 ... 25 ... 100kPa 
Normal range: 0-0.16 ... 0.4 ... 1.0 ... 10 ... 25 ... 60MPa 

Connection: M20*1.5, G1/4, G1/2, 1/4NPT, 1/2NPT 
Accuracy:  0.5%FS, 0.25%FS 

Display Digits:  4 digits LCD 

Dimension: diameter 60mm thickness 38mm 
Power: 12-28VDC( recommend 24VDC) 

Output signal: Rs485 
Output protocol: standard MODBUS RTU 

Communication message: N81/N82/E81/O81 

Sampling rate: 10 times/second 
Output accuracy: 

± 

0.2%FS 

Communication rate: 2400/4800/9600/19200/38400/57600bps 
Power:  <0.4W 
Measurement medium:  water, oil, air etc (non-corrosive medium for SS) 

Transmission distance: 1000 meters 

IP Rating: IP 65 

OPERATION ENVIRONMENT: 
Operation temperature:-20-80

°

Storage temperature:-40-125

°

Environment humidity: 0-95%RH 

1

5. KEY DESCRIPTION

5.1  KEY DESCRIPTION 

Key 

Definition 

Description 

§

Set button 

Lang press to enter the parameter setting state 

@

Short press increase 

lncrease button  When changing the password,  lang press to 

move the location 

® 

Short press to reduce 

Decrease button  When changing the password,  lang press to 

move the location 

5.2  SCREEN  DISPLAY 

�8888

�AR  Kgf/cm'  PSI  MP1  kP� 

5.3  MODE DESCRIPTION 

Status 

Description 

Operation status  Show current pressure and uni! 

Short press lhe 

i§ 

first, then lang 

press@for 

3 seconds 

Zero Calibration  until  the  screen  displays  C--L,  the  zero  clearing  is 

finished (clearing must be performed  without pressing) 

Note:  When the display pressure is higher than 10% 

of the range, it cannot be cleared. 

Setting status 

See the  8th item  for the  general  setting  operation method. 

See the 9th item for how to sei the communication system . 

I

6.DIMENSIONS

____aa_ 

UNIT:  mm 

®J 

UIII

1� 

� 

-

IB) 

[

Radial installation 

Axial installation 

I

7.WIRING METHORDS

RS 485  Digital communication type 

red/brown in 

Digital communication  blue in · 

f24v 

pressure gauge 

black A 

,-,w

=

h

=

ite�B� __ RS 485 

Note: The definition of the outlet is subject to lhe actual label of the product. 

I

8. FUNCTION SETTING

[First set of parameter settings] 

1).Press and hold theif':�button for 2 seconds without releasing until the 

oc 

Loc para!Tleler is displayed. 

2.Click  @  or  ('i")  to get  the  parameter  value,  modify  the  bit  flashing,  lang 

press @a) or(�)to  move  th_e  modified  bit, press  @  or  (�)  to  modify  the 

parameter value,  press  E'.') 

key 

to save. 

3.Change the password lock to 1111  and click the 

E•-� 

button to display the 

next parameter of this group. 

4.Click  the  set  button  to browse  the  parameter  names  in  sequence,  and 
follow  step  @  to  sei  the  parameters  !hat  need  to  be  modified.  When 

reviewing or setting the  last  parameter  of  the  first sei  of  parameters,  click 

lhe sei button to exit lhe setting. 

[The  first parameter setting codes] 

S  b  Name 

Content 

Inner 

ym  O 

add. 

Loc 

Loc 

Password key 

00H 

Rclclc 

Addr  Mailing address 

01H 

bRud 

bAud  Communication 

02H 

rate ootion 

.-dRI: 

rdAt  Communication data  03H 

message forma 

FU:c 

Fltr 

Filter constant 

04H 

Display 

content 

0000-9999 

1-25 

Read & write data 

corresponding value 

0000 H-271 FH 

1 H-FFH 

Note1

OH-SH Note2 

Note3 

0H-3H Note4 

0-20  Note5 

0H-14H 

Отзывы: