D700 | D720 Series Operating Manual
12
pH Buffer Reading vs Temperature
Temp(°C)
pH 4.01
pH 6.86
pH 9.18
pH 4.00
pH 7.00
pH 10.01
0
4.01
6.98
9.47
4.01
7.12
10.32
5
4.01
6.95
9.38
4.00
7.09
10.25
10
4.00
6.92
9.32
4.00
7.06
10.18
15
4.00
6.90
9.27
4.00
7.04
10.12
20
4.00
6.88
9.22
4.00
7.02
10.06
25
4.01
6.86
9.18
4.00
7.00
10.01
30
4.01
6.85
9.14
4.01
6.99
9.97
35
4.02
6.84
9.10
4.02
6.98
9.93
40
4.03
6.84
9.07
4.03
6.97
9.89
45
4.04
6.83
9.04
4.04
6.97
9.86
50
4.06
6.83
9.01
4.06
6.97
9.83
55
4.08
6.83
8.99
4.07
6.97
9.91
60
4.10
6.84
8.96
4.09
6.98
9.97
70
4.12
6.85
8.92
4.12
6.99
9.76
80
4.16
6.86
8.89
4.16
7.00
9.74
90
4.01
6.88
8.85
4.20
7.02
9.73
ORP Calibration:
Press the
[MENU] button to enter the calibration mode, select ORP standard liquid calibration, and input the known
standard liquid value (Default 235mV).
Press the
[MENU] button to return to the measurement screen and press the
[CAL] button to enter the standard solutioncalibration mode. If there is no need to modify the value, this setting is omitted.
Press the
[CAL] button to enter the calibration mode. After entering the calibration mode, the instrument will be displayed as the upper
right image, and the corresponding mV value will be displayed on the left side of the screen.
After the calibration is completed, the slope will be displayed on the right side of the screen. If the instruemnt has been calibrated, press the
[CAL] button to inspect the calibration state, and then press the
[ENT] button to enter the re-calibration.
Field Calibration
Содержание ProCon D700 Series
Страница 1: ...D700 D720 Series pH ORP Digital Analyzer Controller Operating Manual ...
Страница 14: ...D700 D720 Series Operating Manual 14 ...
Страница 15: ...D700 D720 Series Operating Manual 15 Measurement Settings ...
Страница 16: ...D700 D720 Series Operating Manual 16 SimplCal Calibration ...
Страница 17: ...D700 D720 Series Operating Manual 17 Field Calibration ...
Страница 18: ...D700 D720 Series Operating Manual 18 Calibration Adjusment ...
Страница 19: ...D700 D720 Series Operating Manual 19 Alarm Relay Settings ...
Страница 20: ...D700 D720 Series Operating Manual 20 Output Settings ...