background image

Page 6 

6 Bias Adjustment  

& Valve Change 

Read these notes all the way through first. 
If  you  are  unsure  about  any  aspect  contact  your 
retailer, Icon Audio or a competent service engineer. 

 

When  checking  the  bias  ensure  that  any  “Active” 
pre-

amp  or  source  is  “off”  with  zero  volume  to 

prevent  false  readings.  Passive  pre-amps  should 
have “zero” volume. On a test bench “short  circuit” 
the input. 

The MB 90s use 

the “fixed bias’ mode of valve operation. 

This  has  the  advantage  of  higher  power,  and  cooler 
running.  However  occasionally  it  is  advisable  to  check 
the  bias  reading  using  the  built  in  meter  to  ensure  best 
performance  from  the  amplifier.  This  is  very  easy 
involving the flick of a switch. 

 

1,  Tools  you  will  need:

  A  small  flat  blade  screwdriver. 

Run the amplifier for about 10 mins (if possible) first. 

 

2, Reading the bias:

  

Check one amplifier at a time. 
Move  the  Bias  switch  t

o”V1”  the  pointer  in  the  meter 

should be in or near the black section. 

Repeat for “V2”. 

a. If V1 and V2 are about the same but between 55 and 
75, this is normal.  
b. If you find the readings over a few days are generally 
high or low, adjust for the average. 

 

3. How to adjust: 

If one or both valves are high or low, adjust by inserting 
a small screwdriver in each adjuster for the correct valve. 
4. 

The  adjustment  is  very  sensitive  so  adjust  very 

carefully.

  If  the  reading  appears  a  little  unstable  this  is 

normally due to mains fluctuations. 
The  setting  of  one  valve  will  affect  the  other,  so  repeat 
until satisfactory results are obtained. 

 

5. Notes 

If one or more valves are showing erratic readings or you 
cannot  set  the  pointer  in  the  black  section,  then  that 
valve is probably faulty or out of specification. If you are 
unable  to  set  the  reading  high  enough  this  means  the 
emission of the valve is low. 
If  there  is  no  reading  at  all  the  internal  500ma  fuse 
individual  to  each  KT150/120  should  be  checked  by  a 
qualified engineer. Spares should be located inside. 
If the reading is high and cannot be reduced this valve is 
faulty DO NOT USE! 
If the reading is unstable, this usually a valve fault. Swap 
with the other valve to confirm, if the fault moves with the 
valve, suspect the valve.  
If  the  replacement  valve  is  showing  the  same  fault  the 
amplifier may have a fault. 

 

7 Valve Replacement

 

 

Important!

  Do  not  attempt  to  change  the 

KT150/120  output  valves  without  reading  these 
notes.  Failure  to  do  so  could  be  both  dangerous  to 
you and damaging to the amplifier. Keep these notes 
handy. 

Take care that you orient the valve correctly before inserting. Line 
up the centre “Spigot” first. They are easy to break, do not bend 
excessively to the side. 

The  MB90m  MK  II  is  specifically  designed  to  use 
KT120 or KT150 valves. Use of the lower rated KT88, 
6550,  KT90  etc  is  not  recommended  as  the  lower 

ratingings may be  exceeded and will likely result 
in early failure.

 

 

Health & safety; High voltages are present inside 
the amplifier and on exposed valve sockets when 
valves  are  removed,  so  take  suitable  care.  It  is 
not  necessary  to  remove  the  bottom  cover. 
Beware valves get hot in operation! 

 

How do I know when to replace valves? 

 

Generally  speaking  valve  failure  may  be  one  of 
these: 

1,  The  valve  continues  to  work  but  the  emission 
gets low.

 In the case of output valves this will result 

in not being able to set the bias. 

2,  The  valve  gets  noisy/microphonic.

  Usually 

happens  to  the  small  valves,  can  be  confirmed  by 
tapping with a pen. 

3,  Heater  fails.

  No  glow  in  centre  of  valve.  Valve  is 

cold. 
A  valve  that  is  lit  up  is  not  a  guarantee  that  it  is 
working properly; conversely a valve that is not lit up 
will not be working at all. 

4,  Dramatic  Failure.

  Occasionally  the  demise  of  a 

power valve may be obvious with internal sparks and 
noise  through  the  speakers.  In  that  case,  switch  off 
and  do  not  use  until  a  replacement  is  available. 
Before use see chapter 6 “Bias Adjustment”. If there 
was  any  associated  burning  smell  etc,  we  would 
recommend a qualified person examines the inside of 
the amplifier first. 

 

If  the  amplifier  sounds  OK  the  valves  are  probably 
fine.  If  the  emission  drops  you  will  have  difficulty 
setting the bias for the output valves.  

5. Visible Failure: 

Not  all  faults  are  visible,  but  the  following  may  be  a 
sign of failure: 
Heater not lighting, (valve stays cold). 
Valve assembly glowing red, whole or part. 
Blue glow inside the glass. 
Arcing or sparks inside. 
Glass loose on base 
Crack in glass. 
Silver  “getter”  coating  inside  glass  turning  white  or 
being eaten away. 
Excessive rattle when shaken. 
Visible debris inside (often white flecks) inside glass. 

 

6.  Valve  life

  will  depend  upon  such  things  as  hours 

of  use  and  number  of  on/off  cycles,  Use  of  the 
“Standby”  switch  to  warm  the  valves  up  before  use 
will  extend  the  life  of  the  valves.  It  is  not  good 
practice  to  remove  the  valves  unnecessarily  as  this 
can strain the pins and glass and cause tiny air leaks.  

 

7.  Changing  valves:

  If  possible  check  the  bias 

setting before  you  attempt to change the  valve(s), in 
order to familiarise yourself with the procedure.  
If changing all the KT150/120s be ready to adjust the 
bias in order not to overload the power supply. 

Don’t 

worry  how  low  the  reading  goes  this  will  not  cause 
damage.  Do  final  adjustment  when  the  amplifier  is 
fully warmed up. 
If  all  is  well  there  should  be  no  more  than  a  barely 
detectable  hum from  the  speakers,  and  the  amplifier 
should sound OK when tested. 

 

Содержание MB90 MK IIm

Страница 1: ...or feel that we have missed anything out Hi fi reproduction is a long chain of events that includes the recording editing mixing etc before being transferred to a medium such as LP Digital etc before...

Страница 2: ...ck input Output Valve Bias level v Sound Quality Channel Balance Valve Microphony Valve Seating Hum level mv RF Test LED brightness Serial No sticker and recorded Mains voltage 110 230 240V IEC Mains...

Страница 3: ...isible orange glow from the cathode heaters With the pre amp volume control set to minimum there should be no sound coming from the speakers except a barely discernable hum If there are any unpleasant...

Страница 4: ...the speakers close together If correct you should hear plenty of bass if not reverse the terminals for one channel only either at the amp or speaker You will now hear more or less bass The higher bas...

Страница 5: ...ments should be 3 Amp anti surge 5a USA The fuse in the wall plug should be a 3 or 5 amp fuse although unlikely to fail this should be checked if the amplifier fuse is OK No sound Have you selected th...

Страница 6: ...break do not bend excessively to the side The MB90m MK II is specifically designed to use KT120 or KT150 valves Use of the lower rated KT88 6550 KT90 etc is not recommended as the lower ratingings ma...

Страница 7: ...10w RMS KT150 120 Ultralinear 60w RMS KT150 120 Triode mode Signal to noise level 95db Freq response 20Hz 20kHz or 0 1db Power bandwidth 3db 10Hz 30kHz 0 1 THD typical 8w 0 015 1w 1khz Custom hand wou...

Страница 8: ...ide the valve cover If the amplifier is stored in the box keep upright Do not store in damp conditions as this will corrode transformer windings etc If the amplifiers are shipped insurance is desirabl...

Страница 9: ...attached to the base If the fuse blows again a VARIAC or variable transformer will be useful to operate the amplifier on reduced power in order the trace the fault or overload Make sure the bias for t...

Отзывы: