34
35
Symptom
Cannot power on
device
Reflection fault
alarm displays
System Fault
Constant leak
warning sound
when scanning
6. Troubleshooting
Probable Cause(s)
Improper battery
installation
Low battery
Scanning beyond the
detector’s range
Background surface
is absorbing laser
or causing a false
reading to occur
Dust on lens
Laser wavelength
drift
Low Battery
Bad battery contact
Charger fault
Solution
Check to see if the
battery is properly
installed, if not,
remove battery and
install correctly
Check, recharge, or
change the battery
Move closer to
target
Change angle to
the target to obtain
a better reflection
surface
Clean lens
Reboot detector
Perform calibration
Contact ICI
Check, recharge, or
change the battery
Take out battery
and put it back in
Reboot charger
Contact ICI
Symptom
No alarm sounds
when gas detected
Charging
indicator flashes
continuously
Probable Cause(s)
Laser wavelength
drift
Scanning too fast
Background
reflection too high
Bad contact of
battery
Charger fault
Solution
Reboot detector
Perform wavelength
calibration
Change angle to
target
Take out battery
and put it back in
Reboot charger
Contact ICI
• Should you have a problem not listed or the solution does not
work, contact ICI for further assistance.