Bluetooth Indicator
Heart Rate (PR)
Signal Strength
Pulse Bar
Oxygen Level (SpO
2
)
Battery Indicator
OX200 User Manual
1. Read the manual carefully before use.
2. The fingertip sensor is sensitive to motion. Keep your hand still while
in use.
3. The fingertip sensor requires a sufficient blood flow to obtain proper
readings. Poor blood circulation can result in inaccurate readings.
If your hands are cold or you have poor circulation, warm up your
hands before attempting to obtain a reading.
4. A tourniquet, blood pressure cuff or other blood flow hindrances may
also result in inaccurate readings.
5. Nail polish and/or acrylic nails may also affect the readings. We
reccommend turning your finger on it's side when in use.
6. Your finger must be clean for a proper reading.
7. The fingertip sensor must be clean for a proper reading.
8. If a reading is difficult to obtain, switch to another finger or to the
other hand.
9. There are a number of other conditions which may lead to an
inaccurate reading including but not limited to recent medical tests
that included an injection of dyes, use of arterial catheters, a weak
pulse, low levels of hemoglobin in the blood, low perfusion (the
quality of your pulse), elevated levels of dysfunctional hemoglobin,
the strength and type of light that you are in while using the pulse
oximeter and the existence of transmitters within certain ranges of
the sensor during use.
10. The sensor will not alert you if your readings are out of normal range.
There is no SpO
2
alarms; it is not for continuous monitoring, as
indicated by the symbol.
11. Explosion hazard: Do not use the sensor in an explosive atmosphere.
1.
Clean the device with a soft cloth dampened with isopropyl
alcohol. Do not pour or spray any liquids onto the sensor. Do
not allow any liquid to enter any openings in the device. Allow
the sensor to dry thoroughly before use.
2.
The rubber inside of the sensor is composed of medical grade
rubber. It is Latex Free, non-toxic, and not harmful to the skin.
3.
The sensor requires no routine calibration or maintenance.
4.
Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
1. Using your nail, easily open the small rubber cover to expose
the micro USB port. Insert the micro USB end (See A) into the
oximeter's charging port
2. Place the USB end (See B) into any
USB power adapter (not included)
or a USB port on your laptop or computer.
3. When charging, the battery icon
on the display will blink. It is fully charged
when the icon stops blinking.
4. The battery can support 2 hours of continuous
use or approximately 400 measurements.
Note:
Make sure the power adapter’s output
voltage is DC5V and output current is ≤1000mAh
Warnings:
1. Keep the oximeter away from young children - small items
such as the USB cable are a choking hazard. 2. Please do not take
a measurement when the device is charging.
1. Open the clamp as shown in the picture above. Using your
thumb and pointer, gently squeeze the back end of the
unit. Place one of your fingers into the opening. Your finger
should be inserted fully.
2. The unit will automatically turn on when the unit is opened
and a finger is inserted. It will automatically begin taking a
reading.
Remember to keep your hands still for the reading. Do not
shake your finger during the test. It is recommended that
you do not move your body while taking a reading.
3. The display will automatically show your pulse rate and
oxygen level readings after a few seconds. Note: Your
pulse rate will not stabilize. It will constantly change as
your pulse rate is based on your heart beats.
4. When there is no signal, a low signal is detected, or a
finger is removed, the unit will power off automatically.
1. Search for the
ichoiceRelax
app in the App store or Google Play
Store. Download the App.
2. Launch the app and fill out the proper fields (You will need a
valid email address).
3. The app will automatically look for a compatible heart rate
sensor within range. You must place the heart rate sensor on
your finger to turn it on. Once connected, the app will load.*
*You cannot access the app without a heart rate sensor. If there
is no signal from the sensor and/or you remove the sensor from
your finger, the app will lock and begin seaching again for a unit
within range.
ichoiceRelax App Features
COACH
– Exercises to help guide your breathing with on-screen
pacers driven by your live heart rate readings.
RELAX
– Various calming videos with relaxing sounds. Follow the
on-screen pacer driven by your live heart rate readings or close
your eyes and allow the sounds to guide you.
SLEEP -
Choose between calming videos with relaxing sounds to
help you fall asleep. Follow the on-screen pacers driven by your
live heart rate readings or close your eyes and allow the sounds to
guide you.
MONITOR -
Optional meters help you take readings via the
fingertip sensor and compare the results in charts and graphs.
Filter results by last 7 days, week, month, and year.
3 Minute Baseline Meter
- Shows your current relaxation state
and generates an inner balance trend based on your calm vs.
stress readings.
30 Second Spot Check
- Easily obtain your heart rate and
oxygen level at any given time.
The ichoiceRelax App uses continuous and accurate heart rate
readings to guide you into a state of relaxation to help reduce your
stress, calm you down and improve health and focus.
The fingertip sensor sends continuous heart rate readings to the
app to help coach and guide you to breathe at your ideal relaxation
breathing rate.
Everyone’s physiology is different and each person will have their
own ideal relaxation breathing rate based on their current mindset.
Your live heart rate readings (obtained by the heart rate sensor) will
help personalize exercises and guide you to breathe at the rate that
is ideal for you to achieve maximum calmness. Our app synchronizes
with your physiology giving you a personalized result.
1. Display: LED display
Low power indication: Battery icon will blink
2. Power Requirement: Lithium battery (250mAh) for power supply
Battery Life: Up to 2 hours (250mAh, 3.7V lithium battery)
Input Voltage: AC100V-240V, Input frequency: 50Hz-60Hz
Output Voltage: DC5V, Output current: 1000mA (Max)
3. Resolution: ±1% for SpO
2
and ±1 BPM for Pulse Rate
4. Measurement Range: Measuring range for SpO
2
: 70-100%
Measuring range for PR: 30-250 BPM
5. Measurement Accuracy:
SpO
2
: 70-100%, ±2 digits; <70%, no definition
PR: 30~99 BPM, ±2 BPM; 100~250 BPM, ±2%
6. Environment Requirement:
Operation Temperature: 41°F~104°F (5°C~40°C)
Storage Temperature: -13°F~158°F (-25°C~70°C )
Ambient Humidity: 15%~93% no condensation in operation;
≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Automatic Power Off: The fingertip sensor will power off automatically
after 8 seconds when the finger is removed.
8. It is recommended that the product be kept in a dry place. A damp
ambient might affect its lifetime and even might damage the product.
9. Declaration: EMC of this product complies with IEC60601-1-2 standard.
The materials which users can come into contact with are non-toxic
and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Inaccurate measurements may be caused by
FACTORS INCLUDING BUT
NOT LIMITED TO:
12. Ethylene oxide sterilizing or immersing the sensors in liquid.
13. Significant levels of dysfunctional hemoglobin (such ascarbonxy-
hemoglobin or methemoglobin).
14. Intravascular dyes such as indocyanine green or methylene blue.
15. SpO
2
measurements may be adversely affected in the presence of
high ambient light such as direct sunlight. In bright light conditions,
cover the sensor area if necessary.
16. Excessive user movement.
17. High-frequency electrosurgical interference and defibrillators.
18. Venous pulsations.
19. Hypotension, severe vasoconstriction, severe anemia or hypothermia.
My display shows
an error message.
My sensor will not
connect to the app
The app reads
"heart rate error"
I cannot navigate
the app
The sensor will not
give a reading.
The readings are
not accurate.
The sensor will not
power on.
Issue
Possible Reason
Compliance
Test
Solution
Electromagnetic Environment - Guidance
1. You're wearing fingernail polish and/or
arcylic nails
2. Your hands are too cold / there is not
enough blood flowing to the fingertip
3. The battery on your sensor is low
4. Too much movement
1. The battery are low
2. The pulse oximeter is damaged
1. There is an internal error with your unit
1. You are not in the correct app
2. Your fingertip sensor is off
3. Your finger is not inside the sensor
4. The sensor is having trouble getting a
reading
5. Your sensor is not within Bluetooth range
6. Your phone's bluetooth is off
1. Contact customer service
1. Make sure you downloaded
ichoiceRelax app.
2. Make sure your finger is inside the
sensor and the unit is powered on.
3. Rub your hands together to generate
heat and get blood flowing to your
fingertips.
4. Make sure the unit and phone are
close proximity
5. Turn your phone's bluetooth on
1. Charge the unit
2. Contact customer service
04222019_v5
Guidance and manufacturer’s declaration – electromagnetic emission
RF emissions
CISPR 11
Group 1
Class B
RF emission
CISPR 11
The pulse oximeter uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The pulse oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the pulse oximeter should assure that it is used in such an
environment.
Distributed by:
ChoiceMMed America Corp.
2558 Pearl Buck Rd: Suite 8A
Bristol, PA 19007
Phone: (215) 874-0458
www.ichoicerelax.com
ChoiceMMed America Corp.
© 2019 ALL RIGHTS RESERVED
1. Keep the fingertip sensor away from young children. Small parts
may be hazardous if swallowed.
Thank you for purchasing the iChoice SMART pulse ox
relaxation coach. With proper use, this system will teach you how to
breathe at your Personal Best Relaxation Breathing Rate to help reduce
stress, calm you down, and improve your health and focus. The fingertip
sensor transmits vital readings in real-time to the free ichoiceRelax app.
Using these live readings, the app will personalize breathing exercises to
help you achieve optimum relaxation.
The smart fingertip sensor can also be used as a pulse oximeter to spot
check your blood oxygen saturation (in % SpO
2
) and your pulse rate.
The SMART pulse ox relaxation coach is for general wellness
and health applications including sports, fitness, and relaxation
management. It is not intended to be used in prevention, diagnosis, or
treatment of any disease.
Box Contents
1. SMART pulse ox relaxation coach
2. User Manual
3. Carrying Pouch
4. Micro USB Charger
A
B
ichoiceRelax App
Phone Compatibility
iOS 12.0 + and Android 5.0+
1. Turn your finger on it's side
2. Rub your hands together to generate
heat and get blood flowing to your
fingertips
3. Charge your unit
4. Make sure your hand is resting and
there is little to no movement
Symbol
Symbol
Definition
Definition
SpO
2
%
PR bpm
SN
IP22
Oxygen Level
Pulse Rate (beats per minute)
Serial Number
Protected against dripping
water
Follow instructions
for use
Conformity to WEEE
Directive
Type BF applied part
Storage temperature and
relative humidity
Battery Indicator
GENERAL DESCRIPTION
PRECAUTIONS FOR USE
MAINTENANCE
TROUBLE SHOOTING
CONNECTING TO APP (optional)
PULSE OXIMETER DISPLAY
GENERAL USE INSTRUCTIONS
CHARGING THE UNIT
TECHNICAL SPECIFICATIONS
WARNINGS
SYMBOL DEFINITIONS
A normal SpO2 reading typically ranges from 95%-99%.
A normal at rest Pulse Rate reading ranges between 60-100
beats per minute.
If you have questions regarding your SpO2 or Pulse Rate readings
we recommend speaking with your healthcare professional.
Note:
1. The pulse bar indicates the adequacy of Pulse Rate and SpO
2
.
If the bar shows less than 30%, readings may be inaccurate.
2. If the screen displays the signal strength symbol (?), the unit
is having trouble getting a reading. Keep your hands still and
retry.